A wrist-worn PPG sensor coupled with a lightweight algorithm can run on a MCU to enable non-invasive and comfortable monitoring, but ensuring robust PPG-based heart-rate monitoring in the presence of motion artifacts is still an open challenge. Recent state-of-the-art algorithms combine PPG and inertial signals to mitigate the effect of motion artifacts. However, these approaches suffer from limited generality. Moreover, their deployment on MCU-based edge nodes has not been investigated. In this work, we tackle both the aforementioned problems by proposing the use of hardware-friendly Temporal Convolutional Networks (TCN) for PPG-based heart estimation. Starting from a single “seed” TCN, we leverage an automatic Neural Architecture Search (NAS) approach to derive a rich family of models. Among them, we obtain a TCN that outperforms the previous state-of-the-art on the largest PPG dataset available (PPGDalia), achieving a Mean Absolute Error (MAE) of just 3.84 Beats Per Minute (BPM). Furthermore, we tested also a set of smaller yet still accurate (MAE of 5.64 - 6.29 BPM) networks that can be deployed on a commercial MCU (STM32L4) which require as few as 5k parameters and reach a latency of 17.1 ms consuming just 0.21 mJ per inference.

Risso M., Burrello A., Pagliari D.J., Benatti S., Macii E., Benini L., et al. (2021). Robust and energy-efficient PPG-based heart-rate monitoring. 345 E 47TH ST, NEW YORK, NY 10017 USA : Institute of Electrical and Electronics Engineers Inc. [10.1109/ISCAS51556.2021.9401282].

Robust and energy-efficient PPG-based heart-rate monitoring

Burrello A.;Benatti S.;Benini L.;
2021

Abstract

A wrist-worn PPG sensor coupled with a lightweight algorithm can run on a MCU to enable non-invasive and comfortable monitoring, but ensuring robust PPG-based heart-rate monitoring in the presence of motion artifacts is still an open challenge. Recent state-of-the-art algorithms combine PPG and inertial signals to mitigate the effect of motion artifacts. However, these approaches suffer from limited generality. Moreover, their deployment on MCU-based edge nodes has not been investigated. In this work, we tackle both the aforementioned problems by proposing the use of hardware-friendly Temporal Convolutional Networks (TCN) for PPG-based heart estimation. Starting from a single “seed” TCN, we leverage an automatic Neural Architecture Search (NAS) approach to derive a rich family of models. Among them, we obtain a TCN that outperforms the previous state-of-the-art on the largest PPG dataset available (PPGDalia), achieving a Mean Absolute Error (MAE) of just 3.84 Beats Per Minute (BPM). Furthermore, we tested also a set of smaller yet still accurate (MAE of 5.64 - 6.29 BPM) networks that can be deployed on a commercial MCU (STM32L4) which require as few as 5k parameters and reach a latency of 17.1 ms consuming just 0.21 mJ per inference.
2021
Proceedings - IEEE International Symposium on Circuits and Systems
1
5
Risso M., Burrello A., Pagliari D.J., Benatti S., Macii E., Benini L., et al. (2021). Robust and energy-efficient PPG-based heart-rate monitoring. 345 E 47TH ST, NEW YORK, NY 10017 USA : Institute of Electrical and Electronics Engineers Inc. [10.1109/ISCAS51556.2021.9401282].
Risso M.; Burrello A.; Pagliari D.J.; Benatti S.; Macii E.; Benini L.; Poncino M.
File in questo prodotto:
Eventuali allegati, non sono esposti

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11585/870186
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 19
  • ???jsp.display-item.citation.isi??? 14
social impact