During the COVID-19 pandemic, a shortage of mechanical ventilators was reported and ventilator sharing between patients was proposed as an ultimate solution. Two lung simulators were ventilated by one anesthesia machine connected through two respiratory circuits and T-pieces. Five different combinations of compliances (30–50 mL × cmH2O−1) and resistances (5–20 cmH2O × L−1 × s−1) were tested. The ventilation setting was: pressure-controlled ventilation, positive end-expiratory pressure 15 cmH2O, inspiratory pressure 10 cmH2O, respiratory rate 20 bpm. Pressures and flows from all the circuit sections have been recorded and analyzed. Simulated patients with equal compliance and resistance received similar ventilation. Compliance reduction from 50 to 30 mL × cmH2O−1 decreased the tidal volume (VT) by 32% (418 ± 49 vs. 285 ± 17 mL). The resistance increase from 5 to 20 cmH2O × L−1 × s−1 decreased VT by 22% (425 ± 69 vs. 331 ± 51 mL). The maximal alveolar pressure was lower at higher compliance and resistance values and decreased linearly with the time constant (r2 = 0.80, p < 0.001). The minimum alveolar pressure ranged from 15.5 ± 0.04 to 16.57 ± 0.04 cmH2O. Cross-flows between the simulated patients have been recorded in all the tested combinations, during both the inspiratory and expiratory phases. The simultaneous ventilation of two patients with one ventilator may be unable to match individual patient’s needs and has a high risk of cross-interference.

Sharing mechanical ventilator: In vitro evaluation of circuit cross-flows and patient interactions / Colombo S.M.; Battistin M.; Carlesso E.; Vivona L.; Carfagna F.; Valsecchi C.; Florio G.; Carenzo L.; Tonetti T.; Ranieri V.M.; Cecconi M.; Pesenti A.; Grasselli G.; Zanella A.. - In: MEMBRANES. - ISSN 2077-0375. - ELETTRONICO. - 11:7(2021), pp. 547.1-547.14. [10.3390/membranes11070547]

Sharing mechanical ventilator: In vitro evaluation of circuit cross-flows and patient interactions

Tonetti T.;Ranieri V. M.;
2021

Abstract

During the COVID-19 pandemic, a shortage of mechanical ventilators was reported and ventilator sharing between patients was proposed as an ultimate solution. Two lung simulators were ventilated by one anesthesia machine connected through two respiratory circuits and T-pieces. Five different combinations of compliances (30–50 mL × cmH2O−1) and resistances (5–20 cmH2O × L−1 × s−1) were tested. The ventilation setting was: pressure-controlled ventilation, positive end-expiratory pressure 15 cmH2O, inspiratory pressure 10 cmH2O, respiratory rate 20 bpm. Pressures and flows from all the circuit sections have been recorded and analyzed. Simulated patients with equal compliance and resistance received similar ventilation. Compliance reduction from 50 to 30 mL × cmH2O−1 decreased the tidal volume (VT) by 32% (418 ± 49 vs. 285 ± 17 mL). The resistance increase from 5 to 20 cmH2O × L−1 × s−1 decreased VT by 22% (425 ± 69 vs. 331 ± 51 mL). The maximal alveolar pressure was lower at higher compliance and resistance values and decreased linearly with the time constant (r2 = 0.80, p < 0.001). The minimum alveolar pressure ranged from 15.5 ± 0.04 to 16.57 ± 0.04 cmH2O. Cross-flows between the simulated patients have been recorded in all the tested combinations, during both the inspiratory and expiratory phases. The simultaneous ventilation of two patients with one ventilator may be unable to match individual patient’s needs and has a high risk of cross-interference.
2021
Sharing mechanical ventilator: In vitro evaluation of circuit cross-flows and patient interactions / Colombo S.M.; Battistin M.; Carlesso E.; Vivona L.; Carfagna F.; Valsecchi C.; Florio G.; Carenzo L.; Tonetti T.; Ranieri V.M.; Cecconi M.; Pesenti A.; Grasselli G.; Zanella A.. - In: MEMBRANES. - ISSN 2077-0375. - ELETTRONICO. - 11:7(2021), pp. 547.1-547.14. [10.3390/membranes11070547]
Colombo S.M.; Battistin M.; Carlesso E.; Vivona L.; Carfagna F.; Valsecchi C.; Florio G.; Carenzo L.; Tonetti T.; Ranieri V.M.; Cecconi M.; Pesenti A.; Grasselli G.; Zanella A.
File in questo prodotto:
File Dimensione Formato  
Colombo - Sharing Mechanical Ventilator.pdf

accesso aperto

Tipo: Versione (PDF) editoriale
Licenza: Licenza per Accesso Aperto. Creative Commons Attribuzione (CCBY)
Dimensione 1.67 MB
Formato Adobe PDF
1.67 MB Adobe PDF Visualizza/Apri
membranes-11-00547-s001.zip

accesso aperto

Tipo: File Supplementare
Licenza: Licenza per accesso libero gratuito
Dimensione 1.91 MB
Formato Zip File
1.91 MB Zip File Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11585/841348
Citazioni
  • ???jsp.display-item.citation.pmc??? 1
  • Scopus 1
  • ???jsp.display-item.citation.isi??? 3
social impact