The objectives of this study were to (a) assess primary stability of a press-fit cup in a simplified acetabular defect model, filled with compacted cancellous bone chips, and (b) to compare the results with primary stability of a press-fit cup combined with two different types of bone graft substitute in the same defect model. A previously developed acetabular test model made of polyurethane foam was used, in which a mainly medial contained defect was implemented. Three test groups (N = 6 each) were prepared: Cancellous bone chips (bone chips), tricalciumphosphate tetrapods + collagen matrix (tetrapods + coll), bioactive glass S53P4 + polyethylene glycol-glycerol matrix (b.a.glass + PEG). Each material was compacted into the acetabulum and a press-fit cup was implanted. The specimens were loaded dynamically in the direction of the maximum resultant force during level walking. Relative motion between cup and test model was assessed with an optical measurement system. At the last load step (3000 N), inducible displacement was highest for bone chips with median [25th percentile; 75th percentile] value of 113 [110; 114] µm and lowest for b.a.glass + PEG with 91 [89; 93] µm. Migration at this load step was highest for b.a.glass + PEG with 868 [845; 936] µm and lowest for tetrapods + coll with 491 [487; 497] µm. The results show a comparable behavior under load of tetrapods + coll and bone chips and suggest that tetrapods + coll could be an attractive alternative to bone chips. However, so far, this was found for one specific defect type and primary stability should be further investigated in additional/more severe defects.

Primary stability of a press-fit cup in combination with impaction grafting in an acetabular defect model / Schierjott R.A.; Hettich G.; Baxmann M.; Morosato F.; Cristofolini L.; Grupp T.M.. - In: JOURNAL OF ORTHOPAEDIC RESEARCH. - ISSN 0736-0266. - STAMPA. - 39:5(2021), pp. 929-940. [10.1002/jor.24810]

Primary stability of a press-fit cup in combination with impaction grafting in an acetabular defect model

Morosato F.;Cristofolini L.;
2021

Abstract

The objectives of this study were to (a) assess primary stability of a press-fit cup in a simplified acetabular defect model, filled with compacted cancellous bone chips, and (b) to compare the results with primary stability of a press-fit cup combined with two different types of bone graft substitute in the same defect model. A previously developed acetabular test model made of polyurethane foam was used, in which a mainly medial contained defect was implemented. Three test groups (N = 6 each) were prepared: Cancellous bone chips (bone chips), tricalciumphosphate tetrapods + collagen matrix (tetrapods + coll), bioactive glass S53P4 + polyethylene glycol-glycerol matrix (b.a.glass + PEG). Each material was compacted into the acetabulum and a press-fit cup was implanted. The specimens were loaded dynamically in the direction of the maximum resultant force during level walking. Relative motion between cup and test model was assessed with an optical measurement system. At the last load step (3000 N), inducible displacement was highest for bone chips with median [25th percentile; 75th percentile] value of 113 [110; 114] µm and lowest for b.a.glass + PEG with 91 [89; 93] µm. Migration at this load step was highest for b.a.glass + PEG with 868 [845; 936] µm and lowest for tetrapods + coll with 491 [487; 497] µm. The results show a comparable behavior under load of tetrapods + coll and bone chips and suggest that tetrapods + coll could be an attractive alternative to bone chips. However, so far, this was found for one specific defect type and primary stability should be further investigated in additional/more severe defects.
2021
Primary stability of a press-fit cup in combination with impaction grafting in an acetabular defect model / Schierjott R.A.; Hettich G.; Baxmann M.; Morosato F.; Cristofolini L.; Grupp T.M.. - In: JOURNAL OF ORTHOPAEDIC RESEARCH. - ISSN 0736-0266. - STAMPA. - 39:5(2021), pp. 929-940. [10.1002/jor.24810]
Schierjott R.A.; Hettich G.; Baxmann M.; Morosato F.; Cristofolini L.; Grupp T.M.
File in questo prodotto:
File Dimensione Formato  
Schierjott-bone_graft-PU_models-J_Orthop_Res-Open_Access.pdf

accesso aperto

Tipo: Versione (PDF) editoriale
Licenza: Licenza per Accesso Aperto. Creative Commons Attribuzione - Non commerciale (CCBYNC)
Dimensione 3.27 MB
Formato Adobe PDF
3.27 MB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11585/819725
Citazioni
  • ???jsp.display-item.citation.pmc??? 1
  • Scopus 0
  • ???jsp.display-item.citation.isi??? 0
social impact