Objective: The target of this methodological evaluation was the feasibility of long-term monitoring of changes in lung conditions by time-difference electrical impedance tomography (tdEIT). In contrast to ventilation monitoring by tdEIT, the monitoring of end-expiratory (EELIC) or end-inspiratory (EILIC) lung impedance change always requires a reference measurement. Approach: To determine the stability of the used Pulmovista 500® EIT system, as a prerequisite it was initially secured on a resistive phantom for 50 h. By comparing the slopes of EELIC for the whole lung area up to 48 h from 36 pigs ventilated at six positive end-expiratory pressure (PEEP) levels from 0 to 18 cmH2O we found a good agreement (range of r2 = 0.93–1.0) between absolute EIT (aEIT) and tdEIT values. This justified the usage of tdEIT with its superior local resolution compared to aEIT for long-term determination of EELIC. Main results: The EELIC was between −0.07 Ωm day−1 at PEEP 4 and −1.04 Ωm day−1 at PEEP 18 cmH2O. The complex local time pattern for EELIC was roughly quantified by the new parameter, centre of end-expiratory change (CoEEC), in equivalence to the established centre of ventilation (CoV). The ventrally located mean of the CoV was fairly constant in the range of 42%–46% of thorax diameter; however, on the contrary, the CoEEC shifted from about 40% to about 75% in the dorsal direction for PEEP levels of 14 and 18 cmH2O. Significance: The observed shifts started earlier for higher PEEP levels. Changes of EELI could be precisely monitored over a period of 48 h by tdEIT on pigs.

Monitoring lung impedance changes during long-term ventilator-induced lung injury ventilation using electrical impedance tomography / Hahn G.; Niewenhuys J.; Just A.; Tonetti T.; Behnemann T.; Rapetti F.; Collino F.; Vasques F.; Maiolo G.; Romitti F.; Gattinoni L.; Quintel M.; Moerer O.. - In: PHYSIOLOGICAL MEASUREMENT. - ISSN 0967-3334. - ELETTRONICO. - 41:9(2020), pp. 095011.1-095011.10. [10.1088/1361-6579/abb1fb]

Monitoring lung impedance changes during long-term ventilator-induced lung injury ventilation using electrical impedance tomography

Tonetti T.;
2020

Abstract

Objective: The target of this methodological evaluation was the feasibility of long-term monitoring of changes in lung conditions by time-difference electrical impedance tomography (tdEIT). In contrast to ventilation monitoring by tdEIT, the monitoring of end-expiratory (EELIC) or end-inspiratory (EILIC) lung impedance change always requires a reference measurement. Approach: To determine the stability of the used Pulmovista 500® EIT system, as a prerequisite it was initially secured on a resistive phantom for 50 h. By comparing the slopes of EELIC for the whole lung area up to 48 h from 36 pigs ventilated at six positive end-expiratory pressure (PEEP) levels from 0 to 18 cmH2O we found a good agreement (range of r2 = 0.93–1.0) between absolute EIT (aEIT) and tdEIT values. This justified the usage of tdEIT with its superior local resolution compared to aEIT for long-term determination of EELIC. Main results: The EELIC was between −0.07 Ωm day−1 at PEEP 4 and −1.04 Ωm day−1 at PEEP 18 cmH2O. The complex local time pattern for EELIC was roughly quantified by the new parameter, centre of end-expiratory change (CoEEC), in equivalence to the established centre of ventilation (CoV). The ventrally located mean of the CoV was fairly constant in the range of 42%–46% of thorax diameter; however, on the contrary, the CoEEC shifted from about 40% to about 75% in the dorsal direction for PEEP levels of 14 and 18 cmH2O. Significance: The observed shifts started earlier for higher PEEP levels. Changes of EELI could be precisely monitored over a period of 48 h by tdEIT on pigs.
2020
Monitoring lung impedance changes during long-term ventilator-induced lung injury ventilation using electrical impedance tomography / Hahn G.; Niewenhuys J.; Just A.; Tonetti T.; Behnemann T.; Rapetti F.; Collino F.; Vasques F.; Maiolo G.; Romitti F.; Gattinoni L.; Quintel M.; Moerer O.. - In: PHYSIOLOGICAL MEASUREMENT. - ISSN 0967-3334. - ELETTRONICO. - 41:9(2020), pp. 095011.1-095011.10. [10.1088/1361-6579/abb1fb]
Hahn G.; Niewenhuys J.; Just A.; Tonetti T.; Behnemann T.; Rapetti F.; Collino F.; Vasques F.; Maiolo G.; Romitti F.; Gattinoni L.; Quintel M.; Moerer O.
File in questo prodotto:
Eventuali allegati, non sono esposti

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11585/778003
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? 0
  • Scopus 8
  • ???jsp.display-item.citation.isi??? 8
social impact