OBJECTIVE: To report and compare the peri-operative outcomes of patients undergoing open (ORC) and robotic-assisted radical cystectomy (RARC) for bladder cancer performed with a radiofrequency seal and cut device (Caiman®). MATERIALS AND METHODS: Data of patients undergoing ORC or RARC between January 2015 and March 2016 at 6 Italian institutions were prospectively recorded and analyzed. Thirty-and 90-day complications were stratified according to the Martin's criteria and graded according to the Clavien-Dindo classification. Data on operative time, blood loss, transfusion rate, complications, and length of stay were evaluated and compared between the ORC and RARC groups. RESULTS: Thirty-three (66%) and 17 (34%) patients were treated with ORC and RARC, respectively. The median age was 72 (64-78) years. Overall operative time was longer in RARC compared to ORC (389 ± 80.1 vs. 242 ± 62.2 min, p < 0.001), while the estimated blood loss during cystectomy was higher after ORC (370 ± 126.8 vs. 243.3 ± 201.6 ml, p = 0.03). The transfusion rate was significantly higher in the ORC compared to RARC (24.2 vs. 5.9%, p = 0.04). Eight (19%) and 7 (16.7%) patients experienced 30- and 90-day post-operative complications, with no significant difference between ORC and RARC. Length of stay was significantly shorter in RARC group (median 7 vs. 14 days, p < 0.001). CONCLUSION: Open and robot-assisted procedures were safely performed by using a new advanced bipolar seal and cut technology (Caiman®). RARC demonstrated to be superior to ORC in terms of bleeding, transfusion rates and length of hospital stay, despite longer operative time.

Borghesi M, S.R. (2019). Peri-Operative Outcomes after Open and Robot-Assisted Radical Cystectomy by Using an Advanced Bipolar Seal and Cut Technology (Caiman®): A Prospective, Comparative, and Multi-Institutional Study. CURRENT UROLOGY, 12(2), 64-69 [10.1159/000489421].

Peri-Operative Outcomes after Open and Robot-Assisted Radical Cystectomy by Using an Advanced Bipolar Seal and Cut Technology (Caiman®): A Prospective, Comparative, and Multi-Institutional Study.

Borghesi M;Schiavina R;Brunocilla E;
2019

Abstract

OBJECTIVE: To report and compare the peri-operative outcomes of patients undergoing open (ORC) and robotic-assisted radical cystectomy (RARC) for bladder cancer performed with a radiofrequency seal and cut device (Caiman®). MATERIALS AND METHODS: Data of patients undergoing ORC or RARC between January 2015 and March 2016 at 6 Italian institutions were prospectively recorded and analyzed. Thirty-and 90-day complications were stratified according to the Martin's criteria and graded according to the Clavien-Dindo classification. Data on operative time, blood loss, transfusion rate, complications, and length of stay were evaluated and compared between the ORC and RARC groups. RESULTS: Thirty-three (66%) and 17 (34%) patients were treated with ORC and RARC, respectively. The median age was 72 (64-78) years. Overall operative time was longer in RARC compared to ORC (389 ± 80.1 vs. 242 ± 62.2 min, p < 0.001), while the estimated blood loss during cystectomy was higher after ORC (370 ± 126.8 vs. 243.3 ± 201.6 ml, p = 0.03). The transfusion rate was significantly higher in the ORC compared to RARC (24.2 vs. 5.9%, p = 0.04). Eight (19%) and 7 (16.7%) patients experienced 30- and 90-day post-operative complications, with no significant difference between ORC and RARC. Length of stay was significantly shorter in RARC group (median 7 vs. 14 days, p < 0.001). CONCLUSION: Open and robot-assisted procedures were safely performed by using a new advanced bipolar seal and cut technology (Caiman®). RARC demonstrated to be superior to ORC in terms of bleeding, transfusion rates and length of hospital stay, despite longer operative time.
2019
Borghesi M, S.R. (2019). Peri-Operative Outcomes after Open and Robot-Assisted Radical Cystectomy by Using an Advanced Bipolar Seal and Cut Technology (Caiman®): A Prospective, Comparative, and Multi-Institutional Study. CURRENT UROLOGY, 12(2), 64-69 [10.1159/000489421].
Borghesi M, Schiavina R, Antonelli A, Buizza C, Celia A, Parma P, De Concilio B, Mengoni F, Romagnoli D, Saraceni G, Brunocilla E, Porreca A
File in questo prodotto:
Eventuali allegati, non sono esposti

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11585/689118
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? 2
  • Scopus 7
  • ???jsp.display-item.citation.isi??? ND
social impact