This work deals with the structural analysis of screw clamps for external fixator devices. Screw clamps are widely used in external fixators as a means to connect the half-pins to the fixator body. The analysis is carried out by both numerical and experimental techniques, based on the case study of a clamp produced by Citieffe (Bologna, Italy). As a preliminary activity, the tribological parameters involved in the screw–clamp interaction have been characterized by means of a mixed finite element analysis and experimental procedure. Then, an assessment of the current design of the clamp has been carried out. A re-design has been proposed, which, based on some targeted geometrical modifications, allows achieving higher strength requirements with the same overall dimensions and type of materials. A higher load-bearing capability for a given size may allow the fixator to be used on a broader population. Finally, a list of good practices for the design of this kind of clamps has been proposed.

Re-design of a uniplanar, monolateral external fixator / De Agostinis, Massimiliano*. - In: PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS. PART H, JOURNAL OF ENGINEERING IN MEDICINE. - ISSN 0954-4119. - STAMPA. - In press:(2018), pp. 1-12. [10.1177/0954411918762021]

Re-design of a uniplanar, monolateral external fixator

De Agostinis, Massimiliano
2018

Abstract

This work deals with the structural analysis of screw clamps for external fixator devices. Screw clamps are widely used in external fixators as a means to connect the half-pins to the fixator body. The analysis is carried out by both numerical and experimental techniques, based on the case study of a clamp produced by Citieffe (Bologna, Italy). As a preliminary activity, the tribological parameters involved in the screw–clamp interaction have been characterized by means of a mixed finite element analysis and experimental procedure. Then, an assessment of the current design of the clamp has been carried out. A re-design has been proposed, which, based on some targeted geometrical modifications, allows achieving higher strength requirements with the same overall dimensions and type of materials. A higher load-bearing capability for a given size may allow the fixator to be used on a broader population. Finally, a list of good practices for the design of this kind of clamps has been proposed.
2018
Re-design of a uniplanar, monolateral external fixator / De Agostinis, Massimiliano*. - In: PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS. PART H, JOURNAL OF ENGINEERING IN MEDICINE. - ISSN 0954-4119. - STAMPA. - In press:(2018), pp. 1-12. [10.1177/0954411918762021]
De Agostinis, Massimiliano*
File in questo prodotto:
Eventuali allegati, non sono esposti

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11585/631616
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? 1
  • Scopus 4
  • ???jsp.display-item.citation.isi??? 3
social impact