In two work packages of the MARK-AGE project, 37 immunological and physiological biomarkers were measured in 3637 serum, plasma or blood samples in five batches during a period of 4 years. The quality of the serum and plasma samples was very good as judged by the low number of biomarker measurements (only 0.2%) that were rejected because of a high hemolysis, icteria or lipemia of the samples. Using quality control samples, day-to-day and batch variations were determined. The mean inter-assay variation of the five batches were all below 8%, with an average inter-assay coefficient of variation of all biomarkers of 4.0%. Also the precision of the measurements was very good, because all measurements were between 90% and 115% of the defined target values. A possible mix-up of samples was determined by comparison of the extreme testosterone levels of men and women. It was concluded that 3% of the sample identification could be mixed-up. Considering the complex procedure from collection to analysis, including preparation, handling, shipment and storage, of the samples in the MARK-AGE project, both the quality of the samples and the quality of the measurements are very good.

Jansen, E., Beekhof, P., Cremers, J., Weinberger, B., Fiegl, S., Toussaint, O., et al. (2015). Quality control data of physiological and immunological biomarkers measured in serum and plasma. MECHANISMS OF AGEING AND DEVELOPMENT, 151, 54-59 [10.1016/j.mad.2015.06.004].

Quality control data of physiological and immunological biomarkers measured in serum and plasma

CAPRI, MIRIAM;FRANCESCHI, CLAUDIO;
2015

Abstract

In two work packages of the MARK-AGE project, 37 immunological and physiological biomarkers were measured in 3637 serum, plasma or blood samples in five batches during a period of 4 years. The quality of the serum and plasma samples was very good as judged by the low number of biomarker measurements (only 0.2%) that were rejected because of a high hemolysis, icteria or lipemia of the samples. Using quality control samples, day-to-day and batch variations were determined. The mean inter-assay variation of the five batches were all below 8%, with an average inter-assay coefficient of variation of all biomarkers of 4.0%. Also the precision of the measurements was very good, because all measurements were between 90% and 115% of the defined target values. A possible mix-up of samples was determined by comparison of the extreme testosterone levels of men and women. It was concluded that 3% of the sample identification could be mixed-up. Considering the complex procedure from collection to analysis, including preparation, handling, shipment and storage, of the samples in the MARK-AGE project, both the quality of the samples and the quality of the measurements are very good.
2015
Jansen, E., Beekhof, P., Cremers, J., Weinberger, B., Fiegl, S., Toussaint, O., et al. (2015). Quality control data of physiological and immunological biomarkers measured in serum and plasma. MECHANISMS OF AGEING AND DEVELOPMENT, 151, 54-59 [10.1016/j.mad.2015.06.004].
Jansen, Eugène; Beekhof, Piet; Cremers, Johannes; Weinberger, Birgit; Fiegl, Simone; Toussaint, Olivier; Bernhard, Jürgen; Gonos, Efstathios; Capri, M...espandi
File in questo prodotto:
File Dimensione Formato  
Jansen2015_MAD.pdf

accesso aperto

Tipo: Versione (PDF) editoriale
Licenza: Licenza per Accesso Aperto. Creative Commons Attribuzione - Non commerciale - Non opere derivate (CCBYNCND)
Dimensione 587.58 kB
Formato Adobe PDF
587.58 kB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11585/521366
Citazioni
  • ???jsp.display-item.citation.pmc??? 7
  • Scopus 15
  • ???jsp.display-item.citation.isi??? 16
social impact