This study assessed: (i) how the magnitude and direction of principal strains vary for different sideways fall loading directions; (ii) how the principal strains for a sideways fall differ from physiological loading directions; (iii) the fracture mechanism during a sideways fall. Eleven human femurs were instrumented with 16 triaxial strain gauges each. The femurs were non-destructively subjected to: (a) six loading configurations covering the range of physiological loading directions; (b) 12 configurations simulating sideways falls. The femurs were eventually fractured in a sideways fall configuration while high-speed cameras recorded the event. When the same force magnitude was applied, strains were significantly larger in a sideways fall than for physiological loading directions (principal compressive strain was 70% larger in a sideways fall). Also the compressive-to-tensile strain ratio was different: for physiological loading the largest compressive strain was only 30% larger than the largest tensile strain; but for the sideways fall, compressive strains were twice as large as the tensile strains. Principal strains during a sideways fall were nearly perpendicular to the direction of principal strains for physiological loading. In the most critical regions (medial part of the head-neck) the direction of principal strain varied by less than 9° between the different physiological loading conditions, whereas it varied by up to 17° between the sideways fall loading conditions. This was associated with a specific fracture mechanism during sideways fall, where failure initiated on the superior-lateral side (compression) followed by later failure of the medially (tension), often exhibiting a two-peak force-displacement curve.

Strain distribution in the proximal Human femur during in vitro simulated sideways fall / Zani, L.; Erani, P.; Grassi, L.; Taddei, F.; Cristofolini, L.. - In: JOURNAL OF BIOMECHANICS. - ISSN 0021-9290. - STAMPA. - 48:10(2015), pp. S0021929015001013.2130-S0021929015001013.2143. [10.1016/j.jbiomech.2015.02.022]

Strain distribution in the proximal Human femur during in vitro simulated sideways fall

CRISTOFOLINI, LUCA
2015

Abstract

This study assessed: (i) how the magnitude and direction of principal strains vary for different sideways fall loading directions; (ii) how the principal strains for a sideways fall differ from physiological loading directions; (iii) the fracture mechanism during a sideways fall. Eleven human femurs were instrumented with 16 triaxial strain gauges each. The femurs were non-destructively subjected to: (a) six loading configurations covering the range of physiological loading directions; (b) 12 configurations simulating sideways falls. The femurs were eventually fractured in a sideways fall configuration while high-speed cameras recorded the event. When the same force magnitude was applied, strains were significantly larger in a sideways fall than for physiological loading directions (principal compressive strain was 70% larger in a sideways fall). Also the compressive-to-tensile strain ratio was different: for physiological loading the largest compressive strain was only 30% larger than the largest tensile strain; but for the sideways fall, compressive strains were twice as large as the tensile strains. Principal strains during a sideways fall were nearly perpendicular to the direction of principal strains for physiological loading. In the most critical regions (medial part of the head-neck) the direction of principal strain varied by less than 9° between the different physiological loading conditions, whereas it varied by up to 17° between the sideways fall loading conditions. This was associated with a specific fracture mechanism during sideways fall, where failure initiated on the superior-lateral side (compression) followed by later failure of the medially (tension), often exhibiting a two-peak force-displacement curve.
2015
Strain distribution in the proximal Human femur during in vitro simulated sideways fall / Zani, L.; Erani, P.; Grassi, L.; Taddei, F.; Cristofolini, L.. - In: JOURNAL OF BIOMECHANICS. - ISSN 0021-9290. - STAMPA. - 48:10(2015), pp. S0021929015001013.2130-S0021929015001013.2143. [10.1016/j.jbiomech.2015.02.022]
Zani, L.; Erani, P.; Grassi, L.; Taddei, F.; Cristofolini, L.
File in questo prodotto:
File Dimensione Formato  
POST_print-Zani-J_Biomech-2015 v2.pdf

Open Access dal 21/03/2017

Tipo: Postprint
Licenza: Licenza per Accesso Aperto. Creative Commons Attribuzione - Non commerciale - Non opere derivate (CCBYNCND)
Dimensione 1.29 MB
Formato Adobe PDF
1.29 MB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11585/519202
Citazioni
  • ???jsp.display-item.citation.pmc??? 14
  • Scopus 46
  • ???jsp.display-item.citation.isi??? 40
social impact