Aortic counterpulsation (IABP) consists in an ECG-controlled forced deflation and inflation of a balloon positioned in the aorta. The device is designed to decrease the ventricular afterload during systole and to increase the coronary driving pressure during diastole. In biomechanical terms, the IABP improves the mechanical matching between the pump and the load, facilitating the transfer of ventricular energy. This paper describes a completely passive aortic counterpulsation solution, with an intra-aortic balloon without a pumping system, designed to improve the mechanical matching between the ventricle and the artery at very low cost and complexity. The only requirement is an external reservoir to amplify the balloon pulsations due to physiologic arterial pressure pulse. Using a cardiovascular simulator and changing the reservoir pressure, a systolic not negligible (7.8 ml) gas volume exchange between the intra-aortic balloon and the reservoir was measured. The same cardiovascular simulator was used to demonstrate an increase in stroke volume in three conditions of progressive ventricular failure, by detecting a change in systolic and diastolic arterial pressures and stroke volume (SV). The maximal arterial pressure always decreased and the diastolic pressure increased. The SV increased up to 7.8%, demonstrating an arterial elastance reduction and better ventricular-aortic mechanical matching and energy transfer. © 2014 Elsevier Ltd. All rights reserved.

Passive aortic counterpulsation: Biomechanical rationale and bench validation / Ivan Corazza; David Bianchini; Emanuela Marcelli; Laura Cercenelli; Romano Zannoli. - In: JOURNAL OF BIOMECHANICS. - ISSN 0021-9290. - STAMPA. - 47:(2014), pp. 1618-1625. [10.1016/j.jbiomech.2014.03.001]

Passive aortic counterpulsation: Biomechanical rationale and bench validation

CORAZZA, IVAN;BIANCHINI, DAVID;MARCELLI, EMANUELA;CERCENELLI, LAURA;ZANNOLI, ROMANO
2014

Abstract

Aortic counterpulsation (IABP) consists in an ECG-controlled forced deflation and inflation of a balloon positioned in the aorta. The device is designed to decrease the ventricular afterload during systole and to increase the coronary driving pressure during diastole. In biomechanical terms, the IABP improves the mechanical matching between the pump and the load, facilitating the transfer of ventricular energy. This paper describes a completely passive aortic counterpulsation solution, with an intra-aortic balloon without a pumping system, designed to improve the mechanical matching between the ventricle and the artery at very low cost and complexity. The only requirement is an external reservoir to amplify the balloon pulsations due to physiologic arterial pressure pulse. Using a cardiovascular simulator and changing the reservoir pressure, a systolic not negligible (7.8 ml) gas volume exchange between the intra-aortic balloon and the reservoir was measured. The same cardiovascular simulator was used to demonstrate an increase in stroke volume in three conditions of progressive ventricular failure, by detecting a change in systolic and diastolic arterial pressures and stroke volume (SV). The maximal arterial pressure always decreased and the diastolic pressure increased. The SV increased up to 7.8%, demonstrating an arterial elastance reduction and better ventricular-aortic mechanical matching and energy transfer. © 2014 Elsevier Ltd. All rights reserved.
2014
Passive aortic counterpulsation: Biomechanical rationale and bench validation / Ivan Corazza; David Bianchini; Emanuela Marcelli; Laura Cercenelli; Romano Zannoli. - In: JOURNAL OF BIOMECHANICS. - ISSN 0021-9290. - STAMPA. - 47:(2014), pp. 1618-1625. [10.1016/j.jbiomech.2014.03.001]
Ivan Corazza; David Bianchini; Emanuela Marcelli; Laura Cercenelli; Romano Zannoli
File in questo prodotto:
Eventuali allegati, non sono esposti

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11585/268693
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? 0
  • Scopus 5
  • ???jsp.display-item.citation.isi??? 4
social impact