Most commercially available foot prostheses are passive ESR feet, which store and release energy to reduce metabolic costs and improve comfort but cannot adjust to varying walking conditions. In contrast, bionic feet adapt to different tasks but are hindered by high weight, power consumption, and cost. This paper presents MyFlex-ζ, an ESR foot with a variable stiffness system, as a compromise between these two categories. MyFlex-ζadjusts stiffness by varying the sagittal-plane distance between two key points, altering force interactions within the prosthesis and affecting overall stiffness. Clinical tests with three transfemoral amputees evaluated stiffness variation across two sessions: the first subjective, where participants assessed stiffness settings during different tasks, and the second biomechanical, measuring performance parameters. Two participants selected different stiffness settings for various tasks, while the third, with limited perception of stiffness changes, showed less distinction in outcomes. Greater sagittal-plane rotation and higher energy absorption were observed in most tasks with more compliant settings, although one participant's results were limited due to selecting close stiffness settings. Overall, these findings suggest MyFlex-ζoffers adaptability and performance improvements over traditional ESR feet. With further actuation and control system development, MyFlex-ζcould mark significant progress in prosthesis technology.
Tabucol, J., Kooiman, V.G.M., Leopaldi, M., Leijendekkers, R., Selleri, G., Mellini, M., et al. (2025). The MyFlex-ζ Foot: A Variable Stiffness ESR Ankle-Foot Prosthesis. IEEE TRANSACTIONS ON NEURAL SYSTEMS AND REHABILITATION ENGINEERING, 33, 653-663 [10.1109/tnsre.2025.3534096].
The MyFlex-ζ Foot: A Variable Stiffness ESR Ankle-Foot Prosthesis
Tabucol, Johnnidel;Leopaldi, Marco
;Selleri, Giacomo;Carloni, Raffaella;Zucchelli, AndreaUltimo
;Brugo, Tommaso M.
2025
Abstract
Most commercially available foot prostheses are passive ESR feet, which store and release energy to reduce metabolic costs and improve comfort but cannot adjust to varying walking conditions. In contrast, bionic feet adapt to different tasks but are hindered by high weight, power consumption, and cost. This paper presents MyFlex-ζ, an ESR foot with a variable stiffness system, as a compromise between these two categories. MyFlex-ζadjusts stiffness by varying the sagittal-plane distance between two key points, altering force interactions within the prosthesis and affecting overall stiffness. Clinical tests with three transfemoral amputees evaluated stiffness variation across two sessions: the first subjective, where participants assessed stiffness settings during different tasks, and the second biomechanical, measuring performance parameters. Two participants selected different stiffness settings for various tasks, while the third, with limited perception of stiffness changes, showed less distinction in outcomes. Greater sagittal-plane rotation and higher energy absorption were observed in most tasks with more compliant settings, although one participant's results were limited due to selecting close stiffness settings. Overall, these findings suggest MyFlex-ζoffers adaptability and performance improvements over traditional ESR feet. With further actuation and control system development, MyFlex-ζcould mark significant progress in prosthesis technology.| File | Dimensione | Formato | |
|---|---|---|---|
|
2025 - The_MyFlex-_Foot_A_Variable_Stiffness_ESR_Ankle-Foot_Prosthesis.pdf
accesso aperto
Descrizione: 2025 - The_MyFlex-_Foot_A_Variable_Stiffness_ESR_Ankle-Foot_Prosthesis.pd
Tipo:
Versione (PDF) editoriale / Version Of Record
Licenza:
Licenza per Accesso Aperto. Creative Commons Attribuzione (CCBY)
Dimensione
8.98 MB
Formato
Adobe PDF
|
8.98 MB | Adobe PDF | Visualizza/Apri |
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.