The aim of this study was to evaluate the reproducibility and stability of left breast positioning during spirometer-guided deep-inspiration breath-hold (DIBH) radiotherapy using an optical surface imaging system (AlignRT). The AlignRT optical tracking system was used to monitor five left-sided breast cancer patients treated using the Active Breathing Coordinator spirometer with DIBH technique. Treatment plans were created using an automated hybrid-VMAT technique on DIBH CTs. A prescribed dose of 60 Gy to the tumor bed and 50 Gy to the breast in 25 fractions was planned. During each treatment session, the antero-posterior (VRT), superior-inferior (LNG), and lateral (LAT) motion of patients was continuously recorded by AlignRT. The intra-breath-hold stability and the intra- and inter-fraction reproducibility were analyzed for all breath-holds and treatment fractions. The dosimetric impact of the residual motion during DIBH was evaluated from the isocenter shifts amplitudes obtained from the 50%, 90%, and 100% cumulative distribution functions of intra-fractional reproducibility. The positional variations of 590 breath-holds as measured by AlignRT were evaluated. The mean intra-breath-hold stability during DIBH was 1.0 +/- 0.4 mm, 2.1 +/- 1.9 mm, and 0.7 +/- 0.5 mm in the VRT, LNG, and LAT directions, with a maximal value of 8.8 mm in LNG direction. Similarly, the mean intra-breath-hold reproducibility was 1.4 +/- 0.8 mm, 1.7 +/- 1.0 mm, and 0.8 +/- 0.5 mm in the VRT, LNG, and LAT directions, with a maximal value of 4.1 mm in LNG direction. Inter-fractional reproducibility showed better reliability, with difference in breathing levels in all fractions of 0.3 mm on average. Based on tolerance limits corresponding to the 90% cumulative distribution level, gating window widths of 1 mm, 2 mm, and 5 mm in the LAT, VRT, and LNG directions were considered an appropriate choice. In conclusion, despite the use of a dedicated spirometer at constant tidal volume, a non-negligible variability of the breast surface position has been reported during breath-holds. The real-time monitoring of breast surface using surface-guided optical technology is strongly recommended.
Cilla, S., Romano, C., Craus, M., Viola, P., Macchia, G., Boccardi, M., et al. (2023). Reproducibility and stability of spirometer-guided deep inspiration breath-hold in left-breast treatments using an optical surface monitoring system. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 24(6), 1-13 [10.1002/acm2.13922].
Reproducibility and stability of spirometer-guided deep inspiration breath-hold in left-breast treatments using an optical surface monitoring system
Buwenge, Milly;Morganti, Alessio G;
2023
Abstract
The aim of this study was to evaluate the reproducibility and stability of left breast positioning during spirometer-guided deep-inspiration breath-hold (DIBH) radiotherapy using an optical surface imaging system (AlignRT). The AlignRT optical tracking system was used to monitor five left-sided breast cancer patients treated using the Active Breathing Coordinator spirometer with DIBH technique. Treatment plans were created using an automated hybrid-VMAT technique on DIBH CTs. A prescribed dose of 60 Gy to the tumor bed and 50 Gy to the breast in 25 fractions was planned. During each treatment session, the antero-posterior (VRT), superior-inferior (LNG), and lateral (LAT) motion of patients was continuously recorded by AlignRT. The intra-breath-hold stability and the intra- and inter-fraction reproducibility were analyzed for all breath-holds and treatment fractions. The dosimetric impact of the residual motion during DIBH was evaluated from the isocenter shifts amplitudes obtained from the 50%, 90%, and 100% cumulative distribution functions of intra-fractional reproducibility. The positional variations of 590 breath-holds as measured by AlignRT were evaluated. The mean intra-breath-hold stability during DIBH was 1.0 +/- 0.4 mm, 2.1 +/- 1.9 mm, and 0.7 +/- 0.5 mm in the VRT, LNG, and LAT directions, with a maximal value of 8.8 mm in LNG direction. Similarly, the mean intra-breath-hold reproducibility was 1.4 +/- 0.8 mm, 1.7 +/- 1.0 mm, and 0.8 +/- 0.5 mm in the VRT, LNG, and LAT directions, with a maximal value of 4.1 mm in LNG direction. Inter-fractional reproducibility showed better reliability, with difference in breathing levels in all fractions of 0.3 mm on average. Based on tolerance limits corresponding to the 90% cumulative distribution level, gating window widths of 1 mm, 2 mm, and 5 mm in the LAT, VRT, and LNG directions were considered an appropriate choice. In conclusion, despite the use of a dedicated spirometer at constant tidal volume, a non-negligible variability of the breast surface position has been reported during breath-holds. The real-time monitoring of breast surface using surface-guided optical technology is strongly recommended.File | Dimensione | Formato | |
---|---|---|---|
ijerph-20-01084-v2.pdf
accesso aperto
Tipo:
Versione (PDF) editoriale
Licenza:
Licenza per Accesso Aperto. Creative Commons Attribuzione (CCBY)
Dimensione
2.02 MB
Formato
Adobe PDF
|
2.02 MB | Adobe PDF | Visualizza/Apri |
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.