Aim: To calibrate eight-polar bioelectrical impedance analysis (BIA) against dual-energy X-ray absorptiometry (DXA) for the assessment of total and appendicular body composition in healthy adults. Research design: A cross-sectional study was carried out. Subjects: Sixty-eight females and 42 males aged 21-82 years participated in the study. Methods: Whole-body fat-free mass (FFM) and appendicular lean tissue mass (LTM) were measured by DXA; resistance (R) of arms, trunk and legs was measured by eight-polar BIA at frequencies of 5, 50, 250 and 500 kHz; whole-body resistance was calculated as the sum R of arms, trunk and legs. Results: The resistance index (RI), i.e. the height2/resistance ratio, was the best predictor of FFM and appendicular LTM. As compared with weight (Wt), RI at 500 kHz explained 35% more variance of FFM (Radj2 = 0.92 vs 0.57), 45% more variance of LTMarm(Radj2 = 0.93 vs 0.48) and 36% more variance of LTMleg (Radj2 = 0.86 vs 0.50) (p < 0.0001 for all). The contribution of age to the unexplained variance of FFM and appendicular LTM was nil or negligible and the RI x sex interactions were either not significant or not important on practical grounds. The percent root mean square error of the estimate was 6% for FFM and 8% for LTMarm and LTMleg. Conclusion: Eight-polar BIA offers accurate estimates of total and appendicular body composition. The attractive hypothesis that eight-polar BIA is influenced minimally by age and sex should be tested on larger samples including younger individuals.
Malavolti M., Mussi C., Poli M., Fantuzzi A.L., Salvioli G., Battistini N., et al. (2003). Cross-calibration of eight-polar bioelectrical impedance analysis versus dual-energy X-ray absorptiometry for the assessment of total and appendicular body composition in healthy subjects aged 21-82 years. ANNALS OF HUMAN BIOLOGY, 30(4), 380-391 [10.1080/0301446031000095211].
Cross-calibration of eight-polar bioelectrical impedance analysis versus dual-energy X-ray absorptiometry for the assessment of total and appendicular body composition in healthy subjects aged 21-82 years
Bedogni G.Ultimo
2003
Abstract
Aim: To calibrate eight-polar bioelectrical impedance analysis (BIA) against dual-energy X-ray absorptiometry (DXA) for the assessment of total and appendicular body composition in healthy adults. Research design: A cross-sectional study was carried out. Subjects: Sixty-eight females and 42 males aged 21-82 years participated in the study. Methods: Whole-body fat-free mass (FFM) and appendicular lean tissue mass (LTM) were measured by DXA; resistance (R) of arms, trunk and legs was measured by eight-polar BIA at frequencies of 5, 50, 250 and 500 kHz; whole-body resistance was calculated as the sum R of arms, trunk and legs. Results: The resistance index (RI), i.e. the height2/resistance ratio, was the best predictor of FFM and appendicular LTM. As compared with weight (Wt), RI at 500 kHz explained 35% more variance of FFM (Radj2 = 0.92 vs 0.57), 45% more variance of LTMarm(Radj2 = 0.93 vs 0.48) and 36% more variance of LTMleg (Radj2 = 0.86 vs 0.50) (p < 0.0001 for all). The contribution of age to the unexplained variance of FFM and appendicular LTM was nil or negligible and the RI x sex interactions were either not significant or not important on practical grounds. The percent root mean square error of the estimate was 6% for FFM and 8% for LTMarm and LTMleg. Conclusion: Eight-polar BIA offers accurate estimates of total and appendicular body composition. The attractive hypothesis that eight-polar BIA is influenced minimally by age and sex should be tested on larger samples including younger individuals.I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.