Purpose: To report outcomes of endovascular repair (EVAR) of infrarenal abdominal aortic aneurysms (AAAs) with currently-available endografts and identify predictors of technical/clinical failure. Materials and Methods: Patients undergoing EVAR between 2012 and 2020 were prospectively collected and retrospectively analyzed. Technical success (TS: no type I-III endoleaks, renal/hypogastric arteries loss, iliac leg occlusion, conversion to open repair and mortality within 24 postoperative hour), proximal neck-related TS (nr-TS: no proximal type I endoleaks, unplanned renal arteries coverage), and 30-day mortality were assessed as early outcomes. Proximal type I endoleak (ELIa), survival and freedom from reinterventions (FFRs) were assessed during follow-up. Uni/multivariate analysis and Cox-regression were used to identified factors associated with early and follow-up outcomes; FFR and survival were assessed by Kaplan-Meier analysis. Results: A total of 710 were included. Technical success and nr-TS were 692 (98%) and 700 (99%), respectively. The presence of >= 2 hostile anatomical infrarenal neck characteristics was associated with technical failure (odds ratio [OR]: 2.4; 95% confidence interval [CI]: 1.3-4.1; p: 0.007). Infrarenal neck angle >90 degrees (OR: 2.88; 95% CI: 9.6-50.3; p: 0.004), barrel shape (OR: 2.33; 95% CI: 11.1-100.3; p: 0.02) or presence of >= 2 hostile anatomical infrarenal neck characteristics (OR: 2.16; 95% CI: 2.5-5.3; p: 0.03) were independent risk factors for neck-related technical failures. Six (0.8%) patients died within 30 postoperative days. Chronic obstructive pulmonary disease (OR: 16; 95% CI: 1.1-218.3; p: 0.04) and urgent repair (OR: 15; 95% CI: 1.8-119.6; p: 0.01) were independent risk factors for 30-day mortality. The mean follow-up was 53 +/- 13 months. There were 12 (1.7%) ELIa during follow-up. Infrarenal neck length <15 mm (hazard ratio [HR]: 2.8; 95% CI: 1.9-9.6; p: 0.005), diameter >28 mm (HR: 2.7; 95% CI: 1.6-9.5; p: 0.006), angle >= 90 degrees (HR: 2.7; 95% CI: 8.3-50.1; p: 0.007), and persistent type II endoleak (HR: 2.9; 95% CI: 1.6-10.1; p: 0.004) were independent risk factors for ELIa. Freedom from reintervention was 91% at 5 years. The ELIa was an independent risk factor for reinterventions during follow-up (HR: 29.5; 95% CI: 1.4-1.6; p<0.001). Survival was 74% at 5 years with 2 cases (0.3%) of late aortic-related mortality. Peripheral arterial occlusive disease (HR: 1.9; 95% CI: 1.4-3.65; p: 0.03), aneurysm diameter >= 65 mm (HR: 2.2; 95% CI: 1.4-3.26; p<0.001), and infrarenal neck length <15 mm (HR: 1.7; 95% CI: 1.2-2.35; p: 0.04) were independent risk factors for mortality during follow-up. Conclusion: Endovascular repair with currently-available endografts has high TS and low 30-day mortality. Survival and FFRs were satisfactory at mid-term. Pre/postoperative risk factors for technical and clinical failure were identified and they should be considered in EVAR indication and postoperative management to reduce complications and improve mid-term outcome. Clinical Impact Pre and postoperative risk factors for technical and clinical EVAR failure can be identified and they should be considered in EVAR indication and postoperative management to reduce complications and improve mid-term outcome.
Gallitto, E., Faggioli, G., Mascoli, C., Goretti, M., Pini, R., Logiacco, A., et al. (2023). Morphological and Clinical Predictors of Early/Follow-up Failure of the Endovascular Infrarenal Abdominal Aneurysm Repair With Currently Available Endografts. JOURNAL OF ENDOVASCULAR THERAPY, -(-), 1-10 [10.1177/15266028231158312].
Morphological and Clinical Predictors of Early/Follow-up Failure of the Endovascular Infrarenal Abdominal Aneurysm Repair With Currently Available Endografts
Gallitto, Enrico;Faggioli, Gianluca;Mascoli, Chiara;Goretti, Martina;Pini, Rodolfo;Logiacco, Antonino;Rocchi, Cristina;Feroldi, Francesca;Caputo, Stefania;Gargiulo, Mauro
2023
Abstract
Purpose: To report outcomes of endovascular repair (EVAR) of infrarenal abdominal aortic aneurysms (AAAs) with currently-available endografts and identify predictors of technical/clinical failure. Materials and Methods: Patients undergoing EVAR between 2012 and 2020 were prospectively collected and retrospectively analyzed. Technical success (TS: no type I-III endoleaks, renal/hypogastric arteries loss, iliac leg occlusion, conversion to open repair and mortality within 24 postoperative hour), proximal neck-related TS (nr-TS: no proximal type I endoleaks, unplanned renal arteries coverage), and 30-day mortality were assessed as early outcomes. Proximal type I endoleak (ELIa), survival and freedom from reinterventions (FFRs) were assessed during follow-up. Uni/multivariate analysis and Cox-regression were used to identified factors associated with early and follow-up outcomes; FFR and survival were assessed by Kaplan-Meier analysis. Results: A total of 710 were included. Technical success and nr-TS were 692 (98%) and 700 (99%), respectively. The presence of >= 2 hostile anatomical infrarenal neck characteristics was associated with technical failure (odds ratio [OR]: 2.4; 95% confidence interval [CI]: 1.3-4.1; p: 0.007). Infrarenal neck angle >90 degrees (OR: 2.88; 95% CI: 9.6-50.3; p: 0.004), barrel shape (OR: 2.33; 95% CI: 11.1-100.3; p: 0.02) or presence of >= 2 hostile anatomical infrarenal neck characteristics (OR: 2.16; 95% CI: 2.5-5.3; p: 0.03) were independent risk factors for neck-related technical failures. Six (0.8%) patients died within 30 postoperative days. Chronic obstructive pulmonary disease (OR: 16; 95% CI: 1.1-218.3; p: 0.04) and urgent repair (OR: 15; 95% CI: 1.8-119.6; p: 0.01) were independent risk factors for 30-day mortality. The mean follow-up was 53 +/- 13 months. There were 12 (1.7%) ELIa during follow-up. Infrarenal neck length <15 mm (hazard ratio [HR]: 2.8; 95% CI: 1.9-9.6; p: 0.005), diameter >28 mm (HR: 2.7; 95% CI: 1.6-9.5; p: 0.006), angle >= 90 degrees (HR: 2.7; 95% CI: 8.3-50.1; p: 0.007), and persistent type II endoleak (HR: 2.9; 95% CI: 1.6-10.1; p: 0.004) were independent risk factors for ELIa. Freedom from reintervention was 91% at 5 years. The ELIa was an independent risk factor for reinterventions during follow-up (HR: 29.5; 95% CI: 1.4-1.6; p<0.001). Survival was 74% at 5 years with 2 cases (0.3%) of late aortic-related mortality. Peripheral arterial occlusive disease (HR: 1.9; 95% CI: 1.4-3.65; p: 0.03), aneurysm diameter >= 65 mm (HR: 2.2; 95% CI: 1.4-3.26; p<0.001), and infrarenal neck length <15 mm (HR: 1.7; 95% CI: 1.2-2.35; p: 0.04) were independent risk factors for mortality during follow-up. Conclusion: Endovascular repair with currently-available endografts has high TS and low 30-day mortality. Survival and FFRs were satisfactory at mid-term. Pre/postoperative risk factors for technical and clinical failure were identified and they should be considered in EVAR indication and postoperative management to reduce complications and improve mid-term outcome. Clinical Impact Pre and postoperative risk factors for technical and clinical EVAR failure can be identified and they should be considered in EVAR indication and postoperative management to reduce complications and improve mid-term outcome.I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.