Long-term endurance of the cement mantle is fundamental for the survival of cemented hip prostheses. Current protocols to characterize bone cements are unsuitable to predict the actual clinical outcome. The aim of this study was to assess if it is possible to rank cement types having diverse clinical outcome by using a simplified in vitro physiological test. Composite femurs were implanted with identical stems (Lubinus-SPII), using different commercial cement types: CMW1 to represent cement with poor clinical outcome; Simplex-P and Cemex-RX to represent cements with a positive clinical outcome. Implanted femurs were subjected to a validated protocol that simulated a demanding but physiological loading spectrum. Inducible micromotions and permanent migrations were recorded throughout the test. After test completion, the cement mantles were sectioned and inspected with dye penetrants to quantify the fatigue-induced cracks. Micromotions did not differ significantly between cement types (possibly because a successful prosthesis was chosen that is very stable in the host bone). Significant differences were observed in terms of cement cracks: CMW1 induced significantly more numerous and larger cracks than Simplex-P and Cemex-RX; no difference was observed between Simplex-P and Cemex-RX. This indicates that this protocol: (a) can discriminate between 'good' and 'bad' cements and (b) yields consistent results when comparable cements are tested. The proposed protocol overcomes the limitations of existing standardized material tests for bone cements. New cements can be assessed in comparison with other cements with known (positive/negative) clinical outcome, tested with the same protocol.

E. Bialoblocka-Juszczyk, L. Cristofolini, P. Erani, M. Viceconti (2010). Effect of long-term physiological activity on the long-term stem stability of cemented hip arthroplasty: in vitro comparison of three commercial bone cements. PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS. PART H, JOURNAL OF ENGINEERING IN MEDICINE, 224(1), 53-65 [10.1243/09544119JEIM641].

Effect of long-term physiological activity on the long-term stem stability of cemented hip arthroplasty: in vitro comparison of three commercial bone cements

BIALOBLOCKA, EWA;CRISTOFOLINI, LUCA;M. Viceconti
2010

Abstract

Long-term endurance of the cement mantle is fundamental for the survival of cemented hip prostheses. Current protocols to characterize bone cements are unsuitable to predict the actual clinical outcome. The aim of this study was to assess if it is possible to rank cement types having diverse clinical outcome by using a simplified in vitro physiological test. Composite femurs were implanted with identical stems (Lubinus-SPII), using different commercial cement types: CMW1 to represent cement with poor clinical outcome; Simplex-P and Cemex-RX to represent cements with a positive clinical outcome. Implanted femurs were subjected to a validated protocol that simulated a demanding but physiological loading spectrum. Inducible micromotions and permanent migrations were recorded throughout the test. After test completion, the cement mantles were sectioned and inspected with dye penetrants to quantify the fatigue-induced cracks. Micromotions did not differ significantly between cement types (possibly because a successful prosthesis was chosen that is very stable in the host bone). Significant differences were observed in terms of cement cracks: CMW1 induced significantly more numerous and larger cracks than Simplex-P and Cemex-RX; no difference was observed between Simplex-P and Cemex-RX. This indicates that this protocol: (a) can discriminate between 'good' and 'bad' cements and (b) yields consistent results when comparable cements are tested. The proposed protocol overcomes the limitations of existing standardized material tests for bone cements. New cements can be assessed in comparison with other cements with known (positive/negative) clinical outcome, tested with the same protocol.
2010
E. Bialoblocka-Juszczyk, L. Cristofolini, P. Erani, M. Viceconti (2010). Effect of long-term physiological activity on the long-term stem stability of cemented hip arthroplasty: in vitro comparison of three commercial bone cements. PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS. PART H, JOURNAL OF ENGINEERING IN MEDICINE, 224(1), 53-65 [10.1243/09544119JEIM641].
E. Bialoblocka-Juszczyk; L. Cristofolini; P. Erani; M. Viceconti
File in questo prodotto:
Eventuali allegati, non sono esposti

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11585/95899
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 4
  • ???jsp.display-item.citation.isi??? 3
social impact