In orthopaedic oncology, limb salvage procedures are becoming more frequent thanks to recent major improvements in medical imaging, biomechanical modelling and additive manufacturing. For the pelvis, surgical reconstruction with metal implants after tumor resection remains challenging, because of the complex anatomical structures involved. The aim of the present work is to define a consistent overall procedure to guide surgeons and bioengineers for proper implant design. All relevant steps from medical imaging to an accurate 3D anatomical‐based model are here reported. In detail, the anatomical 3D models include bone shapes from CT on the entire pelvic bone, i.e., including both affected and unaffected sides, and position and extension of the tumor and soft tissues from MRI on the affected side. These models are then registered in space, and an initial shape of the personalized implant for the affected side can be properly designed and dimensioned based on the information from the unaffected side. This reported procedure can be fundamental also for virtual pre‐surgical planning, and the design of patient‐specific cutting guides, which would result is a safe margin for tumor cut. The entire procedure is here shown by describing the results in a single real case.
Durastanti G., Belvedere C., Ruggeri M., Donati D.M., Spazzoli B., Leardini A. (2022). A Pelvic Reconstruction Procedure for Custom‐Made Prosthesis Design of Bone Tumor Surgical Treatments. APPLIED SCIENCES, 12(3), 1-12 [10.3390/app12031654].
A Pelvic Reconstruction Procedure for Custom‐Made Prosthesis Design of Bone Tumor Surgical Treatments
Durastanti G.;Belvedere C.;Donati D. M.;Spazzoli B.;Leardini A.
2022
Abstract
In orthopaedic oncology, limb salvage procedures are becoming more frequent thanks to recent major improvements in medical imaging, biomechanical modelling and additive manufacturing. For the pelvis, surgical reconstruction with metal implants after tumor resection remains challenging, because of the complex anatomical structures involved. The aim of the present work is to define a consistent overall procedure to guide surgeons and bioengineers for proper implant design. All relevant steps from medical imaging to an accurate 3D anatomical‐based model are here reported. In detail, the anatomical 3D models include bone shapes from CT on the entire pelvic bone, i.e., including both affected and unaffected sides, and position and extension of the tumor and soft tissues from MRI on the affected side. These models are then registered in space, and an initial shape of the personalized implant for the affected side can be properly designed and dimensioned based on the information from the unaffected side. This reported procedure can be fundamental also for virtual pre‐surgical planning, and the design of patient‐specific cutting guides, which would result is a safe margin for tumor cut. The entire procedure is here shown by describing the results in a single real case.File | Dimensione | Formato | |
---|---|---|---|
applsci-12-01654-v2.pdf
accesso aperto
Tipo:
Versione (PDF) editoriale
Licenza:
Licenza per Accesso Aperto. Creative Commons Attribuzione (CCBY)
Dimensione
3.47 MB
Formato
Adobe PDF
|
3.47 MB | Adobe PDF | Visualizza/Apri |
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.