In this contribution we discuss the calibration and performance of the TOF-Wall detector of the Fragmentation-Of-Target (FOOT) experiment. A thin carbon target was irradiated with carbon ions of several energies ranging from 150 to 400 MeV/u. The charge spectra of the produced fragments were used to calibrate the energy response. The energy resolution for a single bar was about 6%. The time resolution between the two layers ranged from 42 (150 MeV/u) to 62 ps (400 MeV/u).
Kraan A.C., Battistoni G., Belcari N., Bisogni M.G., Carra P., Ciarrocchi E., et al. (2023). Calibration and performance assessment of the TOF-Wall detector of the FOOT experiment. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH. SECTION A, ACCELERATORS, SPECTROMETERS, DETECTORS AND ASSOCIATED EQUIPMENT, 1045, 1-3 [10.1016/j.nima.2022.167615].
Calibration and performance assessment of the TOF-Wall detector of the FOOT experiment
Zarrella R.;
2023
Abstract
In this contribution we discuss the calibration and performance of the TOF-Wall detector of the Fragmentation-Of-Target (FOOT) experiment. A thin carbon target was irradiated with carbon ions of several energies ranging from 150 to 400 MeV/u. The charge spectra of the produced fragments were used to calibrate the energy response. The energy resolution for a single bar was about 6%. The time resolution between the two layers ranged from 42 (150 MeV/u) to 62 ps (400 MeV/u).File | Dimensione | Formato | |
---|---|---|---|
AM.pdf
Open Access dal 21/10/2024
Tipo:
Postprint
Licenza:
Licenza per Accesso Aperto. Creative Commons Attribuzione - Non commerciale - Non opere derivate (CCBYNCND)
Dimensione
1.08 MB
Formato
Adobe PDF
|
1.08 MB | Adobe PDF | Visualizza/Apri |
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.