This work proposes an alternative method to address the correct calibration procedure of Pulsed Electro-Acoustic measurements, which minimizes the possibility of an incorrect evaluation regarding space charge accumulation at the very beginning of polarization in a material. This is done by monitoring the potential of a floating mass in the thickness of the material to be tested. Results show that the proposed method is a viable option, with much less uncertainties than more traditional methods.
Method for the determination of correct calibration pulses in PEA measurements / Mariani D.; Suraci S.V.; Fabiani D.; Seri P.. - ELETTRONICO. - (2022), pp. 1-4. (Intervento presentato al convegno 2022 IEEE International Conference on High Voltage Engineering and Applications, ICHVE 2022 tenutosi a Chongqing, China nel 2022) [10.1109/ICHVE53725.2022.10014482].
Method for the determination of correct calibration pulses in PEA measurements
Mariani D.
Visualization
;Suraci S. V.Validation
;Fabiani D.Supervision
;Seri P.Writing – Original Draft Preparation
2022
Abstract
This work proposes an alternative method to address the correct calibration procedure of Pulsed Electro-Acoustic measurements, which minimizes the possibility of an incorrect evaluation regarding space charge accumulation at the very beginning of polarization in a material. This is done by monitoring the potential of a floating mass in the thickness of the material to be tested. Results show that the proposed method is a viable option, with much less uncertainties than more traditional methods.I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.