This work proposes an alternative method to address the correct calibration procedure of Pulsed Electro-Acoustic measurements, which minimizes the possibility of an incorrect evaluation regarding space charge accumulation at the very beginning of polarization in a material. This is done by monitoring the potential of a floating mass in the thickness of the material to be tested. Results show that the proposed method is a viable option, with much less uncertainties than more traditional methods.
Method for the determination of correct calibration pulses in PEA measurements
Mariani D.
Visualization
;Suraci S. V.Validation
;Fabiani D.Supervision
;Seri P.Writing – Original Draft Preparation
2022
Abstract
This work proposes an alternative method to address the correct calibration procedure of Pulsed Electro-Acoustic measurements, which minimizes the possibility of an incorrect evaluation regarding space charge accumulation at the very beginning of polarization in a material. This is done by monitoring the potential of a floating mass in the thickness of the material to be tested. Results show that the proposed method is a viable option, with much less uncertainties than more traditional methods.File in questo prodotto: