Subclinical valve thrombosis in heart valve prostheses is characterized by the progressive reduction in leaflet motion detectable with advanced imaging diagnostics. However, without routine imaging surveillance, this subclinical thrombosis may be underdiagnosed. We recently proposed the novel concept of a sensorized heart valve prosthesis based on electrical impedance measurement (IntraValvular Impedance, IVI) using miniaturized electrodes embedded in the valve structure to generate a local electric field that is altered by the cyclic movement of the leaflets. In this study, we investigated the feasibility of the novel IVI-sensing concept applied to biological heart valves (BHVs). Three proof-of-concept prototypes of sensorized BHVs were assembled with different size, geometry and positioning of the electrodes to identify the optimal IVI-measurement configuration. Each prototype was tested in vitro on a hydrodynamic heart valve assessment platform. IVI signal was closely related to the electrodes’ positioning in the valve structure and showed greater sensitivity in the prototype with small electrodes embedded in the valve commissures. The novel concept of IVI sensing is feasible on BHVs and has great potential for monitoring the valve condition after implant, allowing for early detection of subclinical valve thrombosis and timely selection of an appropriate anticoagulation therapy.
Gironi, C., Cercenelli, L., Bortolani, B., Emiliani, N., Tartarini, L., Marcelli, E. (2022). Innovative IntraValvular Impedance Sensing Applied to Biological Heart Valve Prostheses: Design and In Vitro Evaluation. SENSORS, 22(21), 1-17 [10.3390/s22218297].
Innovative IntraValvular Impedance Sensing Applied to Biological Heart Valve Prostheses: Design and In Vitro Evaluation
Gironi, CamillaCo-primo
;Cercenelli, Laura
Co-primo
;Bortolani, BarbaraSecondo
;Emiliani, Nicolas;Tartarini, LorenzoPenultimo
;Marcelli, EmanuelaUltimo
2022
Abstract
Subclinical valve thrombosis in heart valve prostheses is characterized by the progressive reduction in leaflet motion detectable with advanced imaging diagnostics. However, without routine imaging surveillance, this subclinical thrombosis may be underdiagnosed. We recently proposed the novel concept of a sensorized heart valve prosthesis based on electrical impedance measurement (IntraValvular Impedance, IVI) using miniaturized electrodes embedded in the valve structure to generate a local electric field that is altered by the cyclic movement of the leaflets. In this study, we investigated the feasibility of the novel IVI-sensing concept applied to biological heart valves (BHVs). Three proof-of-concept prototypes of sensorized BHVs were assembled with different size, geometry and positioning of the electrodes to identify the optimal IVI-measurement configuration. Each prototype was tested in vitro on a hydrodynamic heart valve assessment platform. IVI signal was closely related to the electrodes’ positioning in the valve structure and showed greater sensitivity in the prototype with small electrodes embedded in the valve commissures. The novel concept of IVI sensing is feasible on BHVs and has great potential for monitoring the valve condition after implant, allowing for early detection of subclinical valve thrombosis and timely selection of an appropriate anticoagulation therapy.File | Dimensione | Formato | |
---|---|---|---|
Innovative IntraValvular Impedance Sensing_Gironi_2022.pdf
accesso aperto
Descrizione: Articolo
Tipo:
Versione (PDF) editoriale
Licenza:
Licenza per Accesso Aperto. Creative Commons Attribuzione (CCBY)
Dimensione
7.1 MB
Formato
Adobe PDF
|
7.1 MB | Adobe PDF | Visualizza/Apri |
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.