Liquid chromatography-tandem mass spectrometry (LC-MS/MS) is recommended for measuring circulating steroids. However, assays display technical heterogeneity. So far, reproducibility of corticosteroid LC-MS/MS measurements has received scant attention. The aim of the study was to compare LC-MS/MS measurements of cortisol, 17OH-progesterone and aldosterone from nine European centers and assess performance according to external quality assessment (EQA) materials and calibration. Seventy-eight patient samples, EQA materials and two commercial calibration sets were measured twice by laboratory-specific procedures. Results were obtained by in-house (CAL1) and external calibrations (CAL2 and CAL3). We evaluated intra and inter-laboratory imprecision, correlation and agreement in patient samples, and trueness, bias and commutability in EQA materials. Using CAL1, intra-laboratory CVs ranged between 2.8-7.4%, 4.4-18.0% and 5.2-22.2%, for cortisol, 17OH-progesterone and aldosterone, respectively. Trueness and bias in EQA materials were mostly acceptable, however, inappropriate commutability and target value assignment were highlighted in some cases. CAL2 showed suboptimal accuracy. Median inter-laboratory CVs for cortisol, 17OH-progesterone and aldosterone were 4.9, 11.8 and 13.8% with CAL1 and 3.6, 10.3 and 8.6% with CAL3 (all p<0.001), respectively. Using CAL1, median bias vs. all laboratory-medians ranged from -6.6 to 6.9%, -17.2 to 7.8% and -12.0 to 16.8% for cortisol, 17OH-progesterone and aldosterone, respectively. Regression lines significantly deviated from the best fit for most laboratories. Using CAL3 improved cortisol and 17OH-progesterone between-method bias and correlation. Intra-laboratory imprecision and performance with EQA materials were variable. Inter-laboratory performance was mostly within specifications. Although residual variability persists, adopting common traceable calibrators and RMP-determined EQA materials is beneficial for standardization of LC-MS/MS steroid measurements.
Fanelli, F., Cantu, M., Temchenko, A., Mezzullo, M., Lindner, J.M., Peitzsch, M., et al. (2022). Report from the HarmoSter study: Impact of calibration on comparability of LC-MS/MS measurement of circulating cortisol, 17OH-progesterone and aldosterone. CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 60(0), 726-739 [10.1515/cclm-2021-1028].
Report from the HarmoSter study: Impact of calibration on comparability of LC-MS/MS measurement of circulating cortisol, 17OH-progesterone and aldosterone
Fanelli F.;Temchenko A.;Mezzullo M.;Nardi E.;Pagotto U.
2022
Abstract
Liquid chromatography-tandem mass spectrometry (LC-MS/MS) is recommended for measuring circulating steroids. However, assays display technical heterogeneity. So far, reproducibility of corticosteroid LC-MS/MS measurements has received scant attention. The aim of the study was to compare LC-MS/MS measurements of cortisol, 17OH-progesterone and aldosterone from nine European centers and assess performance according to external quality assessment (EQA) materials and calibration. Seventy-eight patient samples, EQA materials and two commercial calibration sets were measured twice by laboratory-specific procedures. Results were obtained by in-house (CAL1) and external calibrations (CAL2 and CAL3). We evaluated intra and inter-laboratory imprecision, correlation and agreement in patient samples, and trueness, bias and commutability in EQA materials. Using CAL1, intra-laboratory CVs ranged between 2.8-7.4%, 4.4-18.0% and 5.2-22.2%, for cortisol, 17OH-progesterone and aldosterone, respectively. Trueness and bias in EQA materials were mostly acceptable, however, inappropriate commutability and target value assignment were highlighted in some cases. CAL2 showed suboptimal accuracy. Median inter-laboratory CVs for cortisol, 17OH-progesterone and aldosterone were 4.9, 11.8 and 13.8% with CAL1 and 3.6, 10.3 and 8.6% with CAL3 (all p<0.001), respectively. Using CAL1, median bias vs. all laboratory-medians ranged from -6.6 to 6.9%, -17.2 to 7.8% and -12.0 to 16.8% for cortisol, 17OH-progesterone and aldosterone, respectively. Regression lines significantly deviated from the best fit for most laboratories. Using CAL3 improved cortisol and 17OH-progesterone between-method bias and correlation. Intra-laboratory imprecision and performance with EQA materials were variable. Inter-laboratory performance was mostly within specifications. Although residual variability persists, adopting common traceable calibrators and RMP-determined EQA materials is beneficial for standardization of LC-MS/MS steroid measurements.File | Dimensione | Formato | |
---|---|---|---|
Fanelli et al CCLM 2022.pdf
accesso aperto
Tipo:
Versione (PDF) editoriale
Licenza:
Licenza per Accesso Aperto. Creative Commons Attribuzione (CCBY)
Dimensione
759.69 kB
Formato
Adobe PDF
|
759.69 kB | Adobe PDF | Visualizza/Apri |
suppl_j_cclm-2021-1028_suppl.zip
accesso aperto
Tipo:
File Supplementare
Licenza:
Licenza per Accesso Aperto. Altra tipologia di licenza compatibile con Open Access
Dimensione
1.34 MB
Formato
Zip File
|
1.34 MB | Zip File | Visualizza/Apri |
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.