Background and aim: Pulse wave analysis is a pivotal instrument to estimate central hemodynamic parameters. Applanation tonometry on radial and/or carotid arteries is usually used to detect pressure waveforms. Available commercial devices have been validated against invasive catheterism, showing a good agreement of harmonics pattern. In a previous investigation, we observed differences on radial second systolic peak (rSPB2) between two commonly used devices: SphygmoCor (AtCor, Australia) and PulsePen (Diatecne, Italy). The aim of our study was to further quantify differences on radial and carotid signals from the two devices. Methods: We measured radial and carotid pressure waveforms in 38 patients where systolic, diastolic blood pressure and heart rate presented minimal changes between measurements. Waveforms were digitally extracted for off-line analysis. Results: Radial rSBP2, mean arterial pressure, form factor and augmentation index were different with SphygmoCor providing lower values. Carotid augmentation index and form factor were similar. However, carotid systolic pressure (cSBP) from PulsePen was higher that cSBP from SphygmoCor (2.7 ± 4.4. mmHg, P< 0.001). Conclusion: PulsePen and SphygmoCor sensors are not equivalent and provide different wave shapes. These differences on wave shape have important consequences on parameters computed from these waveforms with more discrepancy on radial derived parameters such as rSBP2 and mean arterial pressure than on carotid derived parameters. Further studies are required to compare invasive pressure parameters to indices derived from these two devices. © 2013.

Agnoletti D., Millasseau S., Topouchian J., Zhang Y., Safar M.E., Blacher J. (2013). Pressure wave shape comparison between two non-invasive tonometric devices. ANNALES DE CARDIOLOGIE ET D'ANGEIOLOGIE., 62(3), 193-199 [10.1016/j.ancard.2013.04.002].

Pressure wave shape comparison between two non-invasive tonometric devices

Agnoletti D.
Primo
;
2013

Abstract

Background and aim: Pulse wave analysis is a pivotal instrument to estimate central hemodynamic parameters. Applanation tonometry on radial and/or carotid arteries is usually used to detect pressure waveforms. Available commercial devices have been validated against invasive catheterism, showing a good agreement of harmonics pattern. In a previous investigation, we observed differences on radial second systolic peak (rSPB2) between two commonly used devices: SphygmoCor (AtCor, Australia) and PulsePen (Diatecne, Italy). The aim of our study was to further quantify differences on radial and carotid signals from the two devices. Methods: We measured radial and carotid pressure waveforms in 38 patients where systolic, diastolic blood pressure and heart rate presented minimal changes between measurements. Waveforms were digitally extracted for off-line analysis. Results: Radial rSBP2, mean arterial pressure, form factor and augmentation index were different with SphygmoCor providing lower values. Carotid augmentation index and form factor were similar. However, carotid systolic pressure (cSBP) from PulsePen was higher that cSBP from SphygmoCor (2.7 ± 4.4. mmHg, P< 0.001). Conclusion: PulsePen and SphygmoCor sensors are not equivalent and provide different wave shapes. These differences on wave shape have important consequences on parameters computed from these waveforms with more discrepancy on radial derived parameters such as rSBP2 and mean arterial pressure than on carotid derived parameters. Further studies are required to compare invasive pressure parameters to indices derived from these two devices. © 2013.
2013
Agnoletti D., Millasseau S., Topouchian J., Zhang Y., Safar M.E., Blacher J. (2013). Pressure wave shape comparison between two non-invasive tonometric devices. ANNALES DE CARDIOLOGIE ET D'ANGEIOLOGIE., 62(3), 193-199 [10.1016/j.ancard.2013.04.002].
Agnoletti D.; Millasseau S.; Topouchian J.; Zhang Y.; Safar M.E.; Blacher J.
File in questo prodotto:
Eventuali allegati, non sono esposti

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11585/899554
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 3
  • ???jsp.display-item.citation.isi??? 2
social impact