Background: Four oscillometric devices, including the Omron M6 Comfort, Omron HEM-7420, Withings BP-800, and Polygreen KP-7670, designed for self-blood pressure measurement (SBPM) were evaluated according to the European Society of Hypertension (ESH) International Protocol Revision 2010 in four separate studies. Methods: The four devices measure brachial blood pressure (BP) using the oscillometric method. The Withings BP-800 has to be connected to an Apple® iOS device such as an iPhone®, iPad®, or iPod®. The ESH International Protocol Revision 2010 includes a total number of 33 subjects. The difference between observer and device BP values was calculated for each measure. Ninety-nine pairs of BP differences were classified into three categories (≤5 mmHg, ≤10 mmHg, ≤15 mmHg). The protocol procedures were followed precisely in each of the four studies. Results: All four tested devices passed the validation process. The mean differences between the device and mercury readings were: -1.8±5.1 mmHg and -0.4±2.8 mmHg for systolic and diastolic BP, respectively, using the Omron M6 Comfort device; 2.5±4.6 mmHg and -1.2±4.3 mmHg for the Omron HEM-7420 device; -0.2±5.0 mmHg and 0.4±4.2 mmHg for the Withings BP-800 device; and 3.0±5.3 mmHg and 0.3±5.2 mmHg for the Polygreen KP-7670 device. Conclusion: Omron M6 Comfort, Omron HEM-7420, Withings BP-800, and Polygreen KP-7670 readings differing by less than 5 mmHg, 10 mmHg, and 15 mmHg fulfill the ESH International Protocol Revision 2010 requirements, and therefore are suitable for use by patients for SBPM, if used correctly. © 2014 Topouchian et al.

Validation of four devices: Omron M6 Comfort, Omron HEM-7420, Withings BP-800, and Polygreen KP-7670 for home blood pressure measurement according to the European society of hypertension international protocol / Jirar Topouchian , Davide Agnoletti , Jacques Blacher , Ahmed Youssef , Mirna N Chahine , Isabel Ibanez , Nathalie Assemani , Roland Asmar. - In: VASCULAR HEALTH AND RISK MANAGEMENT. - ISSN 1176-6344. - ELETTRONICO. - 10:(2014), pp. 33-44. [10.2147/VHRM.S53968]

Validation of four devices: Omron M6 Comfort, Omron HEM-7420, Withings BP-800, and Polygreen KP-7670 for home blood pressure measurement according to the European society of hypertension international protocol

Davide Agnoletti
Secondo
;
2014

Abstract

Background: Four oscillometric devices, including the Omron M6 Comfort, Omron HEM-7420, Withings BP-800, and Polygreen KP-7670, designed for self-blood pressure measurement (SBPM) were evaluated according to the European Society of Hypertension (ESH) International Protocol Revision 2010 in four separate studies. Methods: The four devices measure brachial blood pressure (BP) using the oscillometric method. The Withings BP-800 has to be connected to an Apple® iOS device such as an iPhone®, iPad®, or iPod®. The ESH International Protocol Revision 2010 includes a total number of 33 subjects. The difference between observer and device BP values was calculated for each measure. Ninety-nine pairs of BP differences were classified into three categories (≤5 mmHg, ≤10 mmHg, ≤15 mmHg). The protocol procedures were followed precisely in each of the four studies. Results: All four tested devices passed the validation process. The mean differences between the device and mercury readings were: -1.8±5.1 mmHg and -0.4±2.8 mmHg for systolic and diastolic BP, respectively, using the Omron M6 Comfort device; 2.5±4.6 mmHg and -1.2±4.3 mmHg for the Omron HEM-7420 device; -0.2±5.0 mmHg and 0.4±4.2 mmHg for the Withings BP-800 device; and 3.0±5.3 mmHg and 0.3±5.2 mmHg for the Polygreen KP-7670 device. Conclusion: Omron M6 Comfort, Omron HEM-7420, Withings BP-800, and Polygreen KP-7670 readings differing by less than 5 mmHg, 10 mmHg, and 15 mmHg fulfill the ESH International Protocol Revision 2010 requirements, and therefore are suitable for use by patients for SBPM, if used correctly. © 2014 Topouchian et al.
2014
Validation of four devices: Omron M6 Comfort, Omron HEM-7420, Withings BP-800, and Polygreen KP-7670 for home blood pressure measurement according to the European society of hypertension international protocol / Jirar Topouchian , Davide Agnoletti , Jacques Blacher , Ahmed Youssef , Mirna N Chahine , Isabel Ibanez , Nathalie Assemani , Roland Asmar. - In: VASCULAR HEALTH AND RISK MANAGEMENT. - ISSN 1176-6344. - ELETTRONICO. - 10:(2014), pp. 33-44. [10.2147/VHRM.S53968]
Jirar Topouchian , Davide Agnoletti , Jacques Blacher , Ahmed Youssef , Mirna N Chahine , Isabel Ibanez , Nathalie Assemani , Roland Asmar
File in questo prodotto:
Eventuali allegati, non sono esposti

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11585/899542
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? 37
  • Scopus 80
  • ???jsp.display-item.citation.isi??? ND
social impact