Background Although tonometry-based devices have been applied in several population studies and clinical trials, the agreement between them remains unclear. Methods Carotid systolic blood pressure (SBP) and augmentation index (AI) were randomly measured from 3 devices, SphygmoCor, PulsePen, and A-Pulse, in 66 consecutive patients from our ambulatory cardiovascular department. The study contains 2 phases: in Study 1, SphygmoCor and PulsePen were performed on each participant by 2 experienced physicians (n = 66); in Study 2, A-Pulse was added after the measurements of SphygmoCor and PulsePen and performed by another technician on the last 34 patients. Results Carotid SBP and AI measured by the 3 devices were strongly correlated (R ≥ 0.78; P < 0.001), but with significant discrepancies. Specifically, in 66 participants of Study 1, PulsePen estimated higher carotid SBP and AI by 5mm Hg and 5.7%, respectively, than SphygmoCor. In 34 patients of Study 2, A-Pulse estimated higher central SBP by 3.7mm Hg than SphygmoCor, and lower central SBP by 5.7mm Hg than PulsePen. However, no significant difference in interclass comparison was detected between the 3 devices (P ≥ 0.26). Furthermore, slopes of correlation plots of parameters between SphygmoCor and PulsePen were not significantly different from 1 (P ≥ 0.09), but were different in the case of A-Pulse (P ≤ 0.004). Conclusion s Tonometry-based devices were not consistent in measurements of central BP and wave reflections in clinical practice, with considerable and significant differences among them. However, in contrast to A-Pulse, SphygmoCor and PulsePen can probably assess similar cardiovascular risk for individuals, with a systematical discrepancy. © 2012 American Journal of Hypertension, Ltd. All rights reserved.

Zhang Y., Agnoletti D., Safar M.E., Wang J.-G., Topouchian J., Xu Y., et al. (2013). Comparison study of central blood pressure and wave reflection obtained from tonometry-based devices. AMERICAN JOURNAL OF HYPERTENSION, 26(1), 34-41 [10.1093/ajh/hps031].

Comparison study of central blood pressure and wave reflection obtained from tonometry-based devices

Agnoletti D.
Secondo
;
2013

Abstract

Background Although tonometry-based devices have been applied in several population studies and clinical trials, the agreement between them remains unclear. Methods Carotid systolic blood pressure (SBP) and augmentation index (AI) were randomly measured from 3 devices, SphygmoCor, PulsePen, and A-Pulse, in 66 consecutive patients from our ambulatory cardiovascular department. The study contains 2 phases: in Study 1, SphygmoCor and PulsePen were performed on each participant by 2 experienced physicians (n = 66); in Study 2, A-Pulse was added after the measurements of SphygmoCor and PulsePen and performed by another technician on the last 34 patients. Results Carotid SBP and AI measured by the 3 devices were strongly correlated (R ≥ 0.78; P < 0.001), but with significant discrepancies. Specifically, in 66 participants of Study 1, PulsePen estimated higher carotid SBP and AI by 5mm Hg and 5.7%, respectively, than SphygmoCor. In 34 patients of Study 2, A-Pulse estimated higher central SBP by 3.7mm Hg than SphygmoCor, and lower central SBP by 5.7mm Hg than PulsePen. However, no significant difference in interclass comparison was detected between the 3 devices (P ≥ 0.26). Furthermore, slopes of correlation plots of parameters between SphygmoCor and PulsePen were not significantly different from 1 (P ≥ 0.09), but were different in the case of A-Pulse (P ≤ 0.004). Conclusion s Tonometry-based devices were not consistent in measurements of central BP and wave reflections in clinical practice, with considerable and significant differences among them. However, in contrast to A-Pulse, SphygmoCor and PulsePen can probably assess similar cardiovascular risk for individuals, with a systematical discrepancy. © 2012 American Journal of Hypertension, Ltd. All rights reserved.
2013
Zhang Y., Agnoletti D., Safar M.E., Wang J.-G., Topouchian J., Xu Y., et al. (2013). Comparison study of central blood pressure and wave reflection obtained from tonometry-based devices. AMERICAN JOURNAL OF HYPERTENSION, 26(1), 34-41 [10.1093/ajh/hps031].
Zhang Y.; Agnoletti D.; Safar M.E.; Wang J.-G.; Topouchian J.; Xu Y.; Protogerou A.D.; Blacher J.
File in questo prodotto:
Eventuali allegati, non sono esposti

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11585/899533
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 9
  • ???jsp.display-item.citation.isi??? 8
social impact