Objectives: In this study we investigated the rate of susceptibility testing discrepancies between semi-automated and reference systems with carbapenem-resistant Enterobacterales (CRE) and the impact of alleged errors by semi-automated systems on guiding targeted therapy for CRE bloodstream infection (BSI). Methods: This was a multicentre, retrospective study enrolling patients with monomicrobial BSI caused by CRE from January 2013 to December 2016. Nonduplicate isolates from index blood cultures tested locally with semi-automated systems were centralized at a referral laboratory and retested with a reference broth microdilution or agar dilution method. Results: We enrolled 366 patients with CRE-BSI; 220 (60%) were male, and the median age was 67 years (interquartile range, 54–76 years). When compared with the results of the reference methods, those of the semi-automated systems exhibited variable rates of very major errors (VMEs; i.e. false susceptibilities) and major errors (MEs; i.e. false resistances). The highest rates of VMEs were observed with fosfomycin (14%) and colistin (13.9%), and the highest rates of MEs were observed with gentamicin (21%), fosfomycin (7.7%), and tigecycline (34%). Overall, VMEs and MEs led clinicians to prescribe or confirm ineffective therapy in 25 of 341 patients (7%). Receipt of ineffective therapy supported by a misleading susceptibility test was associated with higher 30-day mortality rates by Kaplan–Meier survival curves rates compared with receipt of active therapy (56% vs. 26%; p = 0.002), and the difference was confirmed after adjustment for confounders in a Cox regression model (adjusted hazard ratio: 2.91; 95% CI, 1.62–5.22; p < 0.001). Discussion: MEs and VMEs were relatively common with semi-automated susceptibility testing systems. VMEs were associated with inappropriate use of antibiotics and poorer outcomes.

Bartoletti, M., Antonelli, A., Bussini, L., Corcione, S., Giacobbe, D.R., Marconi, L., et al. (2022). Clinical consequences of very major errors with semi-automated testing systems for antimicrobial susceptibility of carbapenem-resistant Enterobacterales. CLINICAL MICROBIOLOGY AND INFECTION, 28(9), 1290e.1-1290.e4 [10.1016/j.cmi.2022.03.013].

Clinical consequences of very major errors with semi-automated testing systems for antimicrobial susceptibility of carbapenem-resistant Enterobacterales

Bartoletti M.;Bussini L.;Marconi L.;Pascale R.;Ambretti S.;Gaibani P.;Bassetti M.;Lewis R.;Viale P.;Giannella M.
2022

Abstract

Objectives: In this study we investigated the rate of susceptibility testing discrepancies between semi-automated and reference systems with carbapenem-resistant Enterobacterales (CRE) and the impact of alleged errors by semi-automated systems on guiding targeted therapy for CRE bloodstream infection (BSI). Methods: This was a multicentre, retrospective study enrolling patients with monomicrobial BSI caused by CRE from January 2013 to December 2016. Nonduplicate isolates from index blood cultures tested locally with semi-automated systems were centralized at a referral laboratory and retested with a reference broth microdilution or agar dilution method. Results: We enrolled 366 patients with CRE-BSI; 220 (60%) were male, and the median age was 67 years (interquartile range, 54–76 years). When compared with the results of the reference methods, those of the semi-automated systems exhibited variable rates of very major errors (VMEs; i.e. false susceptibilities) and major errors (MEs; i.e. false resistances). The highest rates of VMEs were observed with fosfomycin (14%) and colistin (13.9%), and the highest rates of MEs were observed with gentamicin (21%), fosfomycin (7.7%), and tigecycline (34%). Overall, VMEs and MEs led clinicians to prescribe or confirm ineffective therapy in 25 of 341 patients (7%). Receipt of ineffective therapy supported by a misleading susceptibility test was associated with higher 30-day mortality rates by Kaplan–Meier survival curves rates compared with receipt of active therapy (56% vs. 26%; p = 0.002), and the difference was confirmed after adjustment for confounders in a Cox regression model (adjusted hazard ratio: 2.91; 95% CI, 1.62–5.22; p < 0.001). Discussion: MEs and VMEs were relatively common with semi-automated susceptibility testing systems. VMEs were associated with inappropriate use of antibiotics and poorer outcomes.
2022
Bartoletti, M., Antonelli, A., Bussini, L., Corcione, S., Giacobbe, D.R., Marconi, L., et al. (2022). Clinical consequences of very major errors with semi-automated testing systems for antimicrobial susceptibility of carbapenem-resistant Enterobacterales. CLINICAL MICROBIOLOGY AND INFECTION, 28(9), 1290e.1-1290.e4 [10.1016/j.cmi.2022.03.013].
Bartoletti, M.; Antonelli, A.; Bussini, L.; Corcione, S.; Giacobbe, D. R.; Marconi, L.; Pascale, R.; Dettori, S.; Shbaklo, N.; Ambretti, S.; Gaibani, ...espandi
File in questo prodotto:
Eventuali allegati, non sono esposti

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11585/884328
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? 1
  • Scopus 7
  • ???jsp.display-item.citation.isi??? 10
social impact