A protocol named "Outwalk" was recently proposed to measure the thorax-pelvis and lower-limb kinematics during gait in free-living conditions, by means of an inertial and magnetic measurement system (IMMS). The aim of this study was to validate Outwalk on four healthy subjects when it is used in combination with a specific IMMS (Xsens Technologies, NL), against a reference protocol (CAST) and measurement system (optoelectronic system; Vicon, Oxford Metrics Group, UK). For this purpose, we developed an original approach based on three tests, which allowed to separately investigate: (1) the consequences on joint kinematics of the differences between protocols (Outwalk vs. CAST), (2) the accuracy of the hardware (Xsens vs. Vicon), and (3) the summation of protocols' differences and hardware accuracy (Outwalk + Xsens vs. CAST + Vicon). In order to assess joint-angles similarity, the coefficient of multiple correlation (CMC) was used. For test 3, the CMC showed that Outwalk + Xsens and CAST + Vicon kinematics can be interchanged, offset included, for hip, knee and ankle flexion-extension, and hip ab-adduction (CMC > 0.88). The other joint-angles can be interchanged offset excluded (CMC > 0.85). Tests 1 and 2 also showed that differences in offset between joint-angles were predominantly induced by differences in the protocols; differences in correlation by both hardware and protocols; differences in range of motion by the Xsens accuracy. Results thus support the commencement of a clinical trial of Outwalk on transtibial amputees.

First in vivo assessment of "Outwalk": a novel protocol for clinical gait analysis based on inertial and magnetic sensors / A.Ferrari; A.G.Cutti; P.Garofalo; M.Raggi; M.Heijboer; A.Cappello; A.Davalli. - In: MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING. - ISSN 0140-0118. - STAMPA. - 48:(2010), pp. 1-15. [10.1007/s11517-009-0544-y]

First in vivo assessment of "Outwalk": a novel protocol for clinical gait analysis based on inertial and magnetic sensors

FERRARI, ALBERTO;GAROFALO, PIETRO;CAPPELLO, ANGELO;
2010

Abstract

A protocol named "Outwalk" was recently proposed to measure the thorax-pelvis and lower-limb kinematics during gait in free-living conditions, by means of an inertial and magnetic measurement system (IMMS). The aim of this study was to validate Outwalk on four healthy subjects when it is used in combination with a specific IMMS (Xsens Technologies, NL), against a reference protocol (CAST) and measurement system (optoelectronic system; Vicon, Oxford Metrics Group, UK). For this purpose, we developed an original approach based on three tests, which allowed to separately investigate: (1) the consequences on joint kinematics of the differences between protocols (Outwalk vs. CAST), (2) the accuracy of the hardware (Xsens vs. Vicon), and (3) the summation of protocols' differences and hardware accuracy (Outwalk + Xsens vs. CAST + Vicon). In order to assess joint-angles similarity, the coefficient of multiple correlation (CMC) was used. For test 3, the CMC showed that Outwalk + Xsens and CAST + Vicon kinematics can be interchanged, offset included, for hip, knee and ankle flexion-extension, and hip ab-adduction (CMC > 0.88). The other joint-angles can be interchanged offset excluded (CMC > 0.85). Tests 1 and 2 also showed that differences in offset between joint-angles were predominantly induced by differences in the protocols; differences in correlation by both hardware and protocols; differences in range of motion by the Xsens accuracy. Results thus support the commencement of a clinical trial of Outwalk on transtibial amputees.
2010
First in vivo assessment of "Outwalk": a novel protocol for clinical gait analysis based on inertial and magnetic sensors / A.Ferrari; A.G.Cutti; P.Garofalo; M.Raggi; M.Heijboer; A.Cappello; A.Davalli. - In: MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING. - ISSN 0140-0118. - STAMPA. - 48:(2010), pp. 1-15. [10.1007/s11517-009-0544-y]
A.Ferrari; A.G.Cutti; P.Garofalo; M.Raggi; M.Heijboer; A.Cappello; A.Davalli
File in questo prodotto:
Eventuali allegati, non sono esposti

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11585/86880
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 131
  • ???jsp.display-item.citation.isi??? 122
social impact