Purpose: To evaluate intravascular ultrasound (IVUS) safety and efficacy to detect visceral stenting issues during complex endovascular aneurysm repair through branched and fenestrated repair (B-FEVAR). Materials and Methods: A single-center retrospective analysis of 33 bridging stents assessed intraoperatively using IVUS between January and September 2020 was performed. Ten aortic aneurysm patients [7 thoracoabdominal / 1 pararenal / 2 juxtarenal; 3 females; mean age 73 years [range 70–77 years]) were included. Eight BEVAR (5 standard; 2 custom-made) and 2 FEVAR (custom-made) were performed. The study assessed the safety and efficacy of IVUS utilization to detect immediate branch instability after visceral stenting in the case of B-FEVAR. The primary safety endpoint was defined as the absence of IVUS-related adverse events. The primary efficacy endpoint was defined as the composite of technical success of the IVUS-assessment in each target visceral vessels (TVVs), the rate of IVUS-findings divided as prompting additional maneuvers or not, and the incidence of postoperative computed tomography angiography findings compared with intraoperative assessment. Results: There were no IVUS-related adverse events. The technical success of the IVUS-assessment was achieved in all TVVs. No technical issues compromised the evaluation of the intended vessel. Among the 7 findings identified by IVUS, 3 were suspected at the angiography. In all, 57% (4/7) had normal final angiography. IVUS was able to detect a 12% (4/33) intraoperative branch instability not identified/suspected at the completion angiography. The IVUS assessment led to an immediate revision in 5 (15%) vessels. A total of 57% (4/7) of the issues were detected in patients undergoing primary BEVAR. The remaining 43% (3/7) was detected in patients undergoing secondary intervention for branch instability. Conclusion: IVUS was safe as an adjunctive imaging technique to assess completion after B-FEVAR. It demonstrated efficacy in the detection of intraoperative issues missed by angiography. Further investigations are required to validate these promising results.

Gennai S., Leone N., Saitta G., Migliari M., Lauricella A., Farchioni L., et al. (2021). Intravascular Ultrasound in Branched and Fenestrated Endovascular Aneurysm Repair: Initial Experience in a Single-Center Cohort Study. JOURNAL OF ENDOVASCULAR THERAPY, 28(6), 828-836 [10.1177/15266028211025014].

Intravascular Ultrasound in Branched and Fenestrated Endovascular Aneurysm Repair: Initial Experience in a Single-Center Cohort Study

Leone N.
Co-primo
Writing – Original Draft Preparation
;
Migliari M.;
2021

Abstract

Purpose: To evaluate intravascular ultrasound (IVUS) safety and efficacy to detect visceral stenting issues during complex endovascular aneurysm repair through branched and fenestrated repair (B-FEVAR). Materials and Methods: A single-center retrospective analysis of 33 bridging stents assessed intraoperatively using IVUS between January and September 2020 was performed. Ten aortic aneurysm patients [7 thoracoabdominal / 1 pararenal / 2 juxtarenal; 3 females; mean age 73 years [range 70–77 years]) were included. Eight BEVAR (5 standard; 2 custom-made) and 2 FEVAR (custom-made) were performed. The study assessed the safety and efficacy of IVUS utilization to detect immediate branch instability after visceral stenting in the case of B-FEVAR. The primary safety endpoint was defined as the absence of IVUS-related adverse events. The primary efficacy endpoint was defined as the composite of technical success of the IVUS-assessment in each target visceral vessels (TVVs), the rate of IVUS-findings divided as prompting additional maneuvers or not, and the incidence of postoperative computed tomography angiography findings compared with intraoperative assessment. Results: There were no IVUS-related adverse events. The technical success of the IVUS-assessment was achieved in all TVVs. No technical issues compromised the evaluation of the intended vessel. Among the 7 findings identified by IVUS, 3 were suspected at the angiography. In all, 57% (4/7) had normal final angiography. IVUS was able to detect a 12% (4/33) intraoperative branch instability not identified/suspected at the completion angiography. The IVUS assessment led to an immediate revision in 5 (15%) vessels. A total of 57% (4/7) of the issues were detected in patients undergoing primary BEVAR. The remaining 43% (3/7) was detected in patients undergoing secondary intervention for branch instability. Conclusion: IVUS was safe as an adjunctive imaging technique to assess completion after B-FEVAR. It demonstrated efficacy in the detection of intraoperative issues missed by angiography. Further investigations are required to validate these promising results.
2021
Gennai S., Leone N., Saitta G., Migliari M., Lauricella A., Farchioni L., et al. (2021). Intravascular Ultrasound in Branched and Fenestrated Endovascular Aneurysm Repair: Initial Experience in a Single-Center Cohort Study. JOURNAL OF ENDOVASCULAR THERAPY, 28(6), 828-836 [10.1177/15266028211025014].
Gennai S.; Leone N.; Saitta G.; Migliari M.; Lauricella A.; Farchioni L.; Silingardi R.
File in questo prodotto:
Eventuali allegati, non sono esposti

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11585/852840
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? 0
  • Scopus 12
  • ???jsp.display-item.citation.isi??? 12
social impact