A 57-year-old man with severe aortic valve stenosis at high surgical risk underwent transcatheter valve replacement with a 29-mm Edwards Sapien-3 valve (Edwards Lifesciences). The valve was successfully implanted using the right femoral access, but at the end of the deployment the balloon ruptured (Videos 1 and 2) and could not be retrieved from the e-Sheath. The Commander delivery system was then readvanced in the aorta over the wire (Video 3), which was snared and externalized using a 25-mm AndraSnare (Andramed GmbH) inserted through the left femoral artery using a 25-cm-long 22-F Cook sheath (Cook Medical) (Figure 1). Then, the Commander was pushed over the wire, across the aortic bifurcation, making the ruptured balloon get into the Cook sheath (Figure 2). The Commander was then truncated (Figure 3), and the ruptured balloon proximal part attached to the pusher/balloon catheter unit was retrieved through the eSheath (Figure 4). Then, the ruptured balloon distal part attached to the inner balloon shaft was pushed forward such that it could be externalized through the left femoral sheath, cut (Figure 5), and finally pulled back with the wire through the eSheath (Figure 6).

Bruno, A.G., Taglieri, N., Saia, F., Pini, R., Gallitto, E., Ghetti, G., et al. (2021). Recapture of the Sapien-3 Delivery System After Transversal Balloon Rupture Using a Whole Percutaneous Femoral Approach. JACC: CARDIOVASCULAR INTERVENTIONS, 14(15), 183-187 [10.1016/j.jcin.2021.04.032].

Recapture of the Sapien-3 Delivery System After Transversal Balloon Rupture Using a Whole Percutaneous Femoral Approach

Bruno, Antonio Giulio;Taglieri, Nevio;Saia, Francesco;Pini, Rodolfo;Gallitto, Enrico;Ghetti, Gabriele;Orzalkiewicz, Mateusz;Faggioli, Gianluca;Gargiulo, Mauro;Savini, Carlo;Pacini, Davide;Galié, Nazzareno;Palmerini, Tullio
2021

Abstract

A 57-year-old man with severe aortic valve stenosis at high surgical risk underwent transcatheter valve replacement with a 29-mm Edwards Sapien-3 valve (Edwards Lifesciences). The valve was successfully implanted using the right femoral access, but at the end of the deployment the balloon ruptured (Videos 1 and 2) and could not be retrieved from the e-Sheath. The Commander delivery system was then readvanced in the aorta over the wire (Video 3), which was snared and externalized using a 25-mm AndraSnare (Andramed GmbH) inserted through the left femoral artery using a 25-cm-long 22-F Cook sheath (Cook Medical) (Figure 1). Then, the Commander was pushed over the wire, across the aortic bifurcation, making the ruptured balloon get into the Cook sheath (Figure 2). The Commander was then truncated (Figure 3), and the ruptured balloon proximal part attached to the pusher/balloon catheter unit was retrieved through the eSheath (Figure 4). Then, the ruptured balloon distal part attached to the inner balloon shaft was pushed forward such that it could be externalized through the left femoral sheath, cut (Figure 5), and finally pulled back with the wire through the eSheath (Figure 6).
2021
Bruno, A.G., Taglieri, N., Saia, F., Pini, R., Gallitto, E., Ghetti, G., et al. (2021). Recapture of the Sapien-3 Delivery System After Transversal Balloon Rupture Using a Whole Percutaneous Femoral Approach. JACC: CARDIOVASCULAR INTERVENTIONS, 14(15), 183-187 [10.1016/j.jcin.2021.04.032].
Bruno, Antonio Giulio; Taglieri, Nevio; Saia, Francesco; Pini, Rodolfo; Gallitto, Enrico; Ghetti, Gabriele; Orzalkiewicz, Mateusz; Marrozzini, Cinzia;...espandi
File in questo prodotto:
Eventuali allegati, non sono esposti

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11585/851006
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? 0
  • Scopus 6
  • ???jsp.display-item.citation.isi??? 3
social impact