Purpose: To evaluate the Covera Plus (Bard, Tempe, Arizona) covered stent as a bridging stent in branched thoraco-abdominal endovascular aneurysm repair (BEVAR). Materials and Methods: Retrospective analysis of 112 self-expandable stents was performed at 2 university centers between August 2017 and March 2019. Thirty-one consecutive BEVAR procedures were studied (21 males; mean age, 75 years; 29 American Society of Anesthesiologists grade III/IV) with 119 patent target vessels (7 did not receive the studied device). Primary endpoints were technical success, branch instability, and primary/secondary clinical success. Secondary endpoints included primary/secondary patency and any adverse events. Results: Technical success was 100%. Median follow-up was 12 months (interquartile range, 6–14 months). Branch instability was registered twice (2/112), due to immediate renal occlusion and type Ic endoleak; no additional events occurred during the follow-up. Primary clinical success was 84%. There were 2 immediate deaths, 1 upper limb ischemia, 1 hemorrhage from splenic artery, and 1 type Ic endoleak; these last 3 events were successfully treated, with a secondary clinical success of 100% that was maintained during the follow-up. One late death occurred. Primary patency was 99% (111/112) and 100% (105/105) within 30 days and at the end of the study period,respectively. Adverse events included 4 acute kidney injuries and 2 spinal cord ischemia. No type I/III endoleaks, occlusion, fracture, or dislodging were diagnosed after 30 days. Conclusions: The stent graft seems safe and effective for bridging in BEVAR with low branch instability and high patency rate. Longer follow-up is required to confirm these promising results.

Gennai S., Antonello M., Leone N., Migliari M., Andreoli F., Dall'Antonia A., et al. (2019). Early Experience with the Covera Stent Graft as a Bridging Stent in Branched Thoraco-Abdominal Endovascular Aneurysm Repairs. JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 30(12), 1949-1955 [10.1016/j.jvir.2019.08.006].

Early Experience with the Covera Stent Graft as a Bridging Stent in Branched Thoraco-Abdominal Endovascular Aneurysm Repairs

Leone N.
Writing – Original Draft Preparation
;
Migliari M.;Andreoli F.;
2019

Abstract

Purpose: To evaluate the Covera Plus (Bard, Tempe, Arizona) covered stent as a bridging stent in branched thoraco-abdominal endovascular aneurysm repair (BEVAR). Materials and Methods: Retrospective analysis of 112 self-expandable stents was performed at 2 university centers between August 2017 and March 2019. Thirty-one consecutive BEVAR procedures were studied (21 males; mean age, 75 years; 29 American Society of Anesthesiologists grade III/IV) with 119 patent target vessels (7 did not receive the studied device). Primary endpoints were technical success, branch instability, and primary/secondary clinical success. Secondary endpoints included primary/secondary patency and any adverse events. Results: Technical success was 100%. Median follow-up was 12 months (interquartile range, 6–14 months). Branch instability was registered twice (2/112), due to immediate renal occlusion and type Ic endoleak; no additional events occurred during the follow-up. Primary clinical success was 84%. There were 2 immediate deaths, 1 upper limb ischemia, 1 hemorrhage from splenic artery, and 1 type Ic endoleak; these last 3 events were successfully treated, with a secondary clinical success of 100% that was maintained during the follow-up. One late death occurred. Primary patency was 99% (111/112) and 100% (105/105) within 30 days and at the end of the study period,respectively. Adverse events included 4 acute kidney injuries and 2 spinal cord ischemia. No type I/III endoleaks, occlusion, fracture, or dislodging were diagnosed after 30 days. Conclusions: The stent graft seems safe and effective for bridging in BEVAR with low branch instability and high patency rate. Longer follow-up is required to confirm these promising results.
2019
Gennai S., Antonello M., Leone N., Migliari M., Andreoli F., Dall'Antonia A., et al. (2019). Early Experience with the Covera Stent Graft as a Bridging Stent in Branched Thoraco-Abdominal Endovascular Aneurysm Repairs. JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 30(12), 1949-1955 [10.1016/j.jvir.2019.08.006].
Gennai S.; Antonello M.; Leone N.; Migliari M.; Andreoli F.; Dall'Antonia A.; Grego F.; Silingardi R.
File in questo prodotto:
Eventuali allegati, non sono esposti

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11585/849671
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? 3
  • Scopus 11
  • ???jsp.display-item.citation.isi??? 11
social impact