Background: The subcutaneous implantable cardioverter-defibrillator (S-ICD) does not require the insertion of any leads into the cardiovascular system. Objective: The aims of the study were to describe current practice and to measure outcomes associated with S-ICD or standard single-chamber transvenous ICD (TV-ICD) use after TV-ICD explantation. Methods: We analyzed all consecutive patients who underwent transvenous extraction of an ICD and subsequent implantation of an S-ICD or a single-chamber TV-ICD at 12 Italian centers from 2011 to 2017. Results: A total of 229 patients were extracted and subsequently reimplanted with an S-ICD (90; 39%) or a single-chamber TV-ICD (139; 61%). S-ICD implantation increased from 9% in 2011 to 85% in 2017 (P <.001). Patients reimplanted with an S-ICD were younger (53 ± 13 years vs 60 ± 18 years; P =.011) and more frequently had undergone extraction owing to infection (73% vs 52%; P <.001). The rates of complications at follow-up were comparable between groups (hazard ratio 0.97; 95% confidence interval 0.49–1.92; P =.940). No lead failures, systemic infections, or system-related deaths occurred in the S-ICD group. In the TV-ICD group, 1 lead fracture occurred and 2 systemic infections were reported, resulting in death in 1 case. In the S-ICD group, the rate of complications was lower when the generator was positioned in a sub- or intermuscular pocket (hazard ratio 0.21; 95% confidence interval 0.05–0.87; P =.048). Conclusion: Our results show an increasing use of S-ICD over the years in patients undergoing TV-ICD explantation. An S-ICD is preferably adopted in young patients, mostly in the case of infection. The complication rate was comparable between groups and decreased when a sub- or intermuscular S-ICD generator position was adopted.

Viani S., Migliore F., Tola G., Pisano E.C.L., Russo A.D., Luzzi G., et al. (2019). Use and outcomes of subcutaneous implantable cardioverter-defibrillator (ICD) after transvenous ICD extraction: An analysis of current clinical practice and a comparison with transvenous ICD reimplantation. HEART RHYTHM, 16(4), 564-571 [10.1016/j.hrthm.2018.10.026].

Use and outcomes of subcutaneous implantable cardioverter-defibrillator (ICD) after transvenous ICD extraction: An analysis of current clinical practice and a comparison with transvenous ICD reimplantation

Diemberger I.;
2019

Abstract

Background: The subcutaneous implantable cardioverter-defibrillator (S-ICD) does not require the insertion of any leads into the cardiovascular system. Objective: The aims of the study were to describe current practice and to measure outcomes associated with S-ICD or standard single-chamber transvenous ICD (TV-ICD) use after TV-ICD explantation. Methods: We analyzed all consecutive patients who underwent transvenous extraction of an ICD and subsequent implantation of an S-ICD or a single-chamber TV-ICD at 12 Italian centers from 2011 to 2017. Results: A total of 229 patients were extracted and subsequently reimplanted with an S-ICD (90; 39%) or a single-chamber TV-ICD (139; 61%). S-ICD implantation increased from 9% in 2011 to 85% in 2017 (P <.001). Patients reimplanted with an S-ICD were younger (53 ± 13 years vs 60 ± 18 years; P =.011) and more frequently had undergone extraction owing to infection (73% vs 52%; P <.001). The rates of complications at follow-up were comparable between groups (hazard ratio 0.97; 95% confidence interval 0.49–1.92; P =.940). No lead failures, systemic infections, or system-related deaths occurred in the S-ICD group. In the TV-ICD group, 1 lead fracture occurred and 2 systemic infections were reported, resulting in death in 1 case. In the S-ICD group, the rate of complications was lower when the generator was positioned in a sub- or intermuscular pocket (hazard ratio 0.21; 95% confidence interval 0.05–0.87; P =.048). Conclusion: Our results show an increasing use of S-ICD over the years in patients undergoing TV-ICD explantation. An S-ICD is preferably adopted in young patients, mostly in the case of infection. The complication rate was comparable between groups and decreased when a sub- or intermuscular S-ICD generator position was adopted.
2019
Viani S., Migliore F., Tola G., Pisano E.C.L., Russo A.D., Luzzi G., et al. (2019). Use and outcomes of subcutaneous implantable cardioverter-defibrillator (ICD) after transvenous ICD extraction: An analysis of current clinical practice and a comparison with transvenous ICD reimplantation. HEART RHYTHM, 16(4), 564-571 [10.1016/j.hrthm.2018.10.026].
Viani S.; Migliore F.; Tola G.; Pisano E.C.L.; Russo A.D.; Luzzi G.; Sartori P.; Piro A.; Rordorf R.; Forleo G.B.; Rago A.; Segreti L.; Bertaglia E.; ...espandi
File in questo prodotto:
Eventuali allegati, non sono esposti

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11585/729377
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 36
  • ???jsp.display-item.citation.isi??? 33
social impact