This paper presents a compact, low-cost, and non-invasive system for real-time estimation of plasma sodium concentration ([Na]Pl) during a hemodialysis (HD) session with state-of-the-art accuracy. It is based on electrical impedance spectroscopy (EIS) performed with a capacitively-coupled impedance sensing cell and a high-frequency measurement device, both custom-built. The EIS data are processed to infer the resistance of the liquid inside the cell, which is used together with an optical hemoglobin sensor to estimate the [Na]Pl. Validation of the EIS was performed by estimating the conductivity of blood-mimicking fluid (BMF). The complete method was validated using whole bovine blood, comparing the results to those obtained with standard instruments. The system was able to estimate the [Na]Pl with sufficient accuracy (RMS error of 3.0 mol/m3 with respect to reference data) to provide clinically useful information. The proof-of-concept hardware can be converted to a cheap and
Ravagli, E., Crescentini, M., Rovatti, P., Severi, S. (2020). Non-Invasive Estimation of Plasma Sodium Concentration During Hemodialysis via Capacitively-Coupled Electrical Impedance Spectroscopy. IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 69(4), 1673-1681 [10.1109/TIM.2019.2913612].
Non-Invasive Estimation of Plasma Sodium Concentration During Hemodialysis via Capacitively-Coupled Electrical Impedance Spectroscopy
Ravagli, Enrico;Crescentini, Marco;Severi, Stefano
2020
Abstract
This paper presents a compact, low-cost, and non-invasive system for real-time estimation of plasma sodium concentration ([Na]Pl) during a hemodialysis (HD) session with state-of-the-art accuracy. It is based on electrical impedance spectroscopy (EIS) performed with a capacitively-coupled impedance sensing cell and a high-frequency measurement device, both custom-built. The EIS data are processed to infer the resistance of the liquid inside the cell, which is used together with an optical hemoglobin sensor to estimate the [Na]Pl. Validation of the EIS was performed by estimating the conductivity of blood-mimicking fluid (BMF). The complete method was validated using whole bovine blood, comparing the results to those obtained with standard instruments. The system was able to estimate the [Na]Pl with sufficient accuracy (RMS error of 3.0 mol/m3 with respect to reference data) to provide clinically useful information. The proof-of-concept hardware can be converted to a cheap andI documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.