This paper presents a compact, low-cost, and non-invasive system for real-time estimation of plasma sodium concentration ([Na]Pl) during a hemodialysis (HD) session with state-of-the-art accuracy. It is based on electrical impedance spectroscopy (EIS) performed with a capacitively-coupled impedance sensing cell and a high-frequency measurement device, both custom-built. The EIS data are processed to infer the resistance of the liquid inside the cell, which is used together with an optical hemoglobin sensor to estimate the [Na]Pl. Validation of the EIS was performed by estimating the conductivity of blood-mimicking fluid (BMF). The complete method was validated using whole bovine blood, comparing the results to those obtained with standard instruments. The system was able to estimate the [Na]Pl with sufficient accuracy (RMS error of 3.0 mol/m3 with respect to reference data) to provide clinically useful information. The proof-of-concept hardware can be converted to a cheap and

Non-Invasive Estimation of Plasma Sodium Concentration During Hemodialysis via Capacitively-Coupled Electrical Impedance Spectroscopy / Ravagli, Enrico; Crescentini, Marco; Rovatti, Paolo; Severi, Stefano. - In: IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT. - ISSN 0018-9456. - ELETTRONICO. - 69:4(2020), pp. 1673-1681. [10.1109/TIM.2019.2913612]

Non-Invasive Estimation of Plasma Sodium Concentration During Hemodialysis via Capacitively-Coupled Electrical Impedance Spectroscopy

Ravagli, Enrico;Crescentini, Marco;Severi, Stefano
2020

Abstract

This paper presents a compact, low-cost, and non-invasive system for real-time estimation of plasma sodium concentration ([Na]Pl) during a hemodialysis (HD) session with state-of-the-art accuracy. It is based on electrical impedance spectroscopy (EIS) performed with a capacitively-coupled impedance sensing cell and a high-frequency measurement device, both custom-built. The EIS data are processed to infer the resistance of the liquid inside the cell, which is used together with an optical hemoglobin sensor to estimate the [Na]Pl. Validation of the EIS was performed by estimating the conductivity of blood-mimicking fluid (BMF). The complete method was validated using whole bovine blood, comparing the results to those obtained with standard instruments. The system was able to estimate the [Na]Pl with sufficient accuracy (RMS error of 3.0 mol/m3 with respect to reference data) to provide clinically useful information. The proof-of-concept hardware can be converted to a cheap and
2020
Non-Invasive Estimation of Plasma Sodium Concentration During Hemodialysis via Capacitively-Coupled Electrical Impedance Spectroscopy / Ravagli, Enrico; Crescentini, Marco; Rovatti, Paolo; Severi, Stefano. - In: IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT. - ISSN 0018-9456. - ELETTRONICO. - 69:4(2020), pp. 1673-1681. [10.1109/TIM.2019.2913612]
Ravagli, Enrico; Crescentini, Marco; Rovatti, Paolo; Severi, Stefano
File in questo prodotto:
Eventuali allegati, non sono esposti

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11585/697696
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 12
  • ???jsp.display-item.citation.isi??? 11
social impact