There is significant interest in the stress-strain state in the proximal femoral metaphysis, because of its relevance for hip fractures and prosthetic replacements. The scope of this work was to provide a better understanding of the strain distribution, and of its correlation with the different directions of loading, and with bone quality. A total of 12 pairs of human femurs were instrumented with strain gauges. Six loading configurations were designed to cover the range of directions spanned by the hip joint force. Inter-specimen variability was reduced if paired specimens were considered. The principal strain magnitude varied greatly between loading configurations. This suggests that different loading configurations need to be simulated in vitro. The strain magnitude varied between locations but, on average, was compatible with the strain values measured in vivo. The strain magnitudes and the direction of principal tensile strain in the head and neck were compatible with the spontaneous fractures of the proximal femur reported in some subjects. The principal tensile strain was significantly larger where the cortical bone was thinner; the compressive strain was larger where the cortical bone was thicker. The direction of the principal strain varied significantly between measurement locations but varied little between loading configurations. This suggests that the anatomy and the distribution of anisotropic material properties enable the proximal femur to respond adequately to the changing direction of daily loading.

Strain distribution in the proximal human femoral metaphysis

CRISTOFOLINI, LUCA;JUSZCZYK, MATEUSZ MARIA;TADDEI, FULVIA;VICECONTI, MARCO
2009

Abstract

There is significant interest in the stress-strain state in the proximal femoral metaphysis, because of its relevance for hip fractures and prosthetic replacements. The scope of this work was to provide a better understanding of the strain distribution, and of its correlation with the different directions of loading, and with bone quality. A total of 12 pairs of human femurs were instrumented with strain gauges. Six loading configurations were designed to cover the range of directions spanned by the hip joint force. Inter-specimen variability was reduced if paired specimens were considered. The principal strain magnitude varied greatly between loading configurations. This suggests that different loading configurations need to be simulated in vitro. The strain magnitude varied between locations but, on average, was compatible with the strain values measured in vivo. The strain magnitudes and the direction of principal tensile strain in the head and neck were compatible with the spontaneous fractures of the proximal femur reported in some subjects. The principal tensile strain was significantly larger where the cortical bone was thinner; the compressive strain was larger where the cortical bone was thicker. The direction of the principal strain varied significantly between measurement locations but varied little between loading configurations. This suggests that the anatomy and the distribution of anisotropic material properties enable the proximal femur to respond adequately to the changing direction of daily loading.
File in questo prodotto:
Eventuali allegati, non sono esposti

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11585/66442
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? 6
  • Scopus 58
  • ???jsp.display-item.citation.isi??? 53
social impact