The present study was aimed at investigating the effects of the sterilization method and accelerated aging on the wear and morphology of ultrahigh molecular weight polyethylene. In a first test, gamma- and EtO-sterilized acetabular cups were tested in a hip joint simulator for two million cycles. After the test, the cups underwent an accelerated aging treatment (80 °C, 4 weeks in air furnace) and were newly tested for another three million cycles. Wear was evaluated by gravimetric measurements, morphology by micro-Raman spectroscopy. During the first test, the EtO-sterilized cups underwent a significantly higher wear than the gamma-sterilized ones (62 and 30 mg/million cycles, respectively). No significant crystallinity changes were observed. Upon accelerated aging, the crystallinity increase of the gamma-sterilized cups was more pronounced than for the EtO-sterilized cups, due to chain scission and oxygen incorporation. In the second test, the wear rate of EtO-sterilized cups decreased to 38 mg/million cycles, while for gamma-irradiated cups it increased to 84 mg/million cycles. At the same time, the latter cups underwent significant increases in temperature and crystallinity, due to the higher friction. For the EtO-sterilized cups a significant decrease in crystallinity was observed, due to the occurrence of an orthorhombic → monoclinic phase transformation.
P. Taddei, S. Affatato, M. Rocchi, C. Fagnano, M. Viceconti (2008). The effects of irradiation and EtO-treatment on ultrahigh molecular weight polyethylene acetabular cups following accelerated aging: degradation of mechanical properties and morphology changes during hip simulator tests. JOURNAL OF MOLECULAR STRUCTURE, 875, 254-263 [10.1016/j.molstruc.2007.04.042].
The effects of irradiation and EtO-treatment on ultrahigh molecular weight polyethylene acetabular cups following accelerated aging: degradation of mechanical properties and morphology changes during hip simulator tests.
TADDEI, PAOLA;FAGNANO, CONCEZIO;M. Viceconti
2008
Abstract
The present study was aimed at investigating the effects of the sterilization method and accelerated aging on the wear and morphology of ultrahigh molecular weight polyethylene. In a first test, gamma- and EtO-sterilized acetabular cups were tested in a hip joint simulator for two million cycles. After the test, the cups underwent an accelerated aging treatment (80 °C, 4 weeks in air furnace) and were newly tested for another three million cycles. Wear was evaluated by gravimetric measurements, morphology by micro-Raman spectroscopy. During the first test, the EtO-sterilized cups underwent a significantly higher wear than the gamma-sterilized ones (62 and 30 mg/million cycles, respectively). No significant crystallinity changes were observed. Upon accelerated aging, the crystallinity increase of the gamma-sterilized cups was more pronounced than for the EtO-sterilized cups, due to chain scission and oxygen incorporation. In the second test, the wear rate of EtO-sterilized cups decreased to 38 mg/million cycles, while for gamma-irradiated cups it increased to 84 mg/million cycles. At the same time, the latter cups underwent significant increases in temperature and crystallinity, due to the higher friction. For the EtO-sterilized cups a significant decrease in crystallinity was observed, due to the occurrence of an orthorhombic → monoclinic phase transformation.I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.