Background: Endovascular aortic repair (EVAR) is presently the preferred treatment for abdominal aortic aneurysm; however, it requires the injection of a contrast medium, which can hamper the renal function. Other nontoxic agents such as carbon dioxide (CO2) have been sporadically tested in this setting with uncertain results. The aim of the study is to investigate the efficacy of a new standardized CO2injection method in standard EVAR procedures. Methods: Between August and October 2016, 31 consecutive patients (median age 76.1 [interquartile range IQR: 7.4] years) were submitted to standard EVAR. Proximal and distal endograft landing zones were identified by the injection of 100 mL of CO2at 300 mm Hg, through an 11 cm 10F femoral sheath by a specifically manufactured automated injection device (Angiodroid Srl, San Lazzaro, Bologna, Italy). Before EVAR deployment, a confirmative injection with a conventional contrast medium was accomplished. The possibility of precisely visualizing the proximal and distal landing zones by CO2digital subtraction angiography (DSA) was evaluated considering the contrast medium injection obtained in the same procedure as a gold standard. Similarly, the possible presence of endoleak was assessed at the end of the procedure by the 2 techniques. Results: CO2DSA allowed to identify the juxtarenal landing zone of the endograft in 19/31 cases (61%) and the distal one in 31/31 (100%). In 12 (39%) cases, CO2injection failed to visualize at least the lowest renal artery. This occurred in large aneurysms with scarce thrombotic apposition and a luminal volume greater than 95.9 (IQR: 25.2) mm3. Completion CO2DSA detected type II endoleaks (ELIIs) in 10 cases compared with 2 of conventional contrast media. Conclusions: The injection of nontoxic CO2through an automated device allowed to perform EVAR procedures effectively, in the majority of cases. In some cases, a single injection of a minimum amount of conventional contrast medium can be used to overcome the lack of renal artery visualization by CO2. ELIIs are more frequently visualized with CO2compared with standard contrast medium. Although the CO2injection technique needs further amelioration particularly in the renal arteries detection, this technique appears promising and possibly substitutive of the standard contrast medium, with significant benefit for the renal function.

Standardization of a Carbon Dioxide Automated System for Endovascular Aortic Aneurysm Repair

Mascoli, Chiara;Faggioli, Gianluca;Gallitto, Enrico;Vento, Vincenzo;Pini, Rodolfo;Vacirca, Andrea;Indelicato, Giuseppe;Gargiulo, Mauro;Stella, Andrea
2018

Abstract

Background: Endovascular aortic repair (EVAR) is presently the preferred treatment for abdominal aortic aneurysm; however, it requires the injection of a contrast medium, which can hamper the renal function. Other nontoxic agents such as carbon dioxide (CO2) have been sporadically tested in this setting with uncertain results. The aim of the study is to investigate the efficacy of a new standardized CO2injection method in standard EVAR procedures. Methods: Between August and October 2016, 31 consecutive patients (median age 76.1 [interquartile range IQR: 7.4] years) were submitted to standard EVAR. Proximal and distal endograft landing zones were identified by the injection of 100 mL of CO2at 300 mm Hg, through an 11 cm 10F femoral sheath by a specifically manufactured automated injection device (Angiodroid Srl, San Lazzaro, Bologna, Italy). Before EVAR deployment, a confirmative injection with a conventional contrast medium was accomplished. The possibility of precisely visualizing the proximal and distal landing zones by CO2digital subtraction angiography (DSA) was evaluated considering the contrast medium injection obtained in the same procedure as a gold standard. Similarly, the possible presence of endoleak was assessed at the end of the procedure by the 2 techniques. Results: CO2DSA allowed to identify the juxtarenal landing zone of the endograft in 19/31 cases (61%) and the distal one in 31/31 (100%). In 12 (39%) cases, CO2injection failed to visualize at least the lowest renal artery. This occurred in large aneurysms with scarce thrombotic apposition and a luminal volume greater than 95.9 (IQR: 25.2) mm3. Completion CO2DSA detected type II endoleaks (ELIIs) in 10 cases compared with 2 of conventional contrast media. Conclusions: The injection of nontoxic CO2through an automated device allowed to perform EVAR procedures effectively, in the majority of cases. In some cases, a single injection of a minimum amount of conventional contrast medium can be used to overcome the lack of renal artery visualization by CO2. ELIIs are more frequently visualized with CO2compared with standard contrast medium. Although the CO2injection technique needs further amelioration particularly in the renal arteries detection, this technique appears promising and possibly substitutive of the standard contrast medium, with significant benefit for the renal function.
2018
Mascoli, Chiara; Faggioli, Gianluca*; Gallitto, Enrico; Vento, Vincenzo; Pini, Rodolfo; Vacirca, Andrea; Indelicato, Giuseppe; Gargiulo, Mauro; Stella, Andrea
File in questo prodotto:
Eventuali allegati, non sono esposti

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11585/643844
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? 2
  • Scopus 16
  • ???jsp.display-item.citation.isi??? 15
social impact