Wearable technology for assistive medical applications and physical activity recognition has emerged as a fast growing research field in recent years. However, the design of such systems still poses challenges, including restricted physical size, limited computational resources, and the availability of constrained energy sources. In this paper, we present a practical design space exploration of a body-worn system for prosthetic hand control based on surface electromyography signals (sEMG). The sEMG method is a well-established sensing technology that provides detection of electrical activity produced by the physiological contractions of muscles. The presented wearable system is designed to acquire sEMG signals for successive recognition of the performed motion and the control of the prosthetic hand, allowing to regain a considerable amount of life quality for a broad patient community. The main guiding requirements for the presented wearable system include real-time data acquisition from up to 32 sEMG channels, reliable wireless data streaming, long mobile autonomy (several days lifetime), non-intrusive mounting and compact size.We present the system’s hardware and software architecture focusing on the comparison of various communication design options such as recent Bluetooth Low Energy (BLE) and Low-Power WiFi technologies.

Brunelli, D., Farella, E., Giovanelli, D., Milosevic, B., Minakov, I. (2016). Design Considerations for Wireless Acquisition of Multichannel sEMG Signals in Prosthetic Hand Control. IEEE SENSORS JOURNAL, 16(23), 8338-8347 [10.1109/JSEN.2016.2596712].

Design Considerations for Wireless Acquisition of Multichannel sEMG Signals in Prosthetic Hand Control

BRUNELLI, DAVIDE;FARELLA, ELISABETTA;GIOVANELLI, DAVIDE;MILOSEVIC, BOJAN;
2016

Abstract

Wearable technology for assistive medical applications and physical activity recognition has emerged as a fast growing research field in recent years. However, the design of such systems still poses challenges, including restricted physical size, limited computational resources, and the availability of constrained energy sources. In this paper, we present a practical design space exploration of a body-worn system for prosthetic hand control based on surface electromyography signals (sEMG). The sEMG method is a well-established sensing technology that provides detection of electrical activity produced by the physiological contractions of muscles. The presented wearable system is designed to acquire sEMG signals for successive recognition of the performed motion and the control of the prosthetic hand, allowing to regain a considerable amount of life quality for a broad patient community. The main guiding requirements for the presented wearable system include real-time data acquisition from up to 32 sEMG channels, reliable wireless data streaming, long mobile autonomy (several days lifetime), non-intrusive mounting and compact size.We present the system’s hardware and software architecture focusing on the comparison of various communication design options such as recent Bluetooth Low Energy (BLE) and Low-Power WiFi technologies.
2016
Brunelli, D., Farella, E., Giovanelli, D., Milosevic, B., Minakov, I. (2016). Design Considerations for Wireless Acquisition of Multichannel sEMG Signals in Prosthetic Hand Control. IEEE SENSORS JOURNAL, 16(23), 8338-8347 [10.1109/JSEN.2016.2596712].
Brunelli, Davide; Farella, Elisabetta; Giovanelli, Davide; Milosevic, Bojan; Minakov, Ivan
File in questo prodotto:
Eventuali allegati, non sono esposti

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11585/564738
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 37
  • ???jsp.display-item.citation.isi??? 31
social impact