OBJECTIVE: When handling Y-labelled and Lu-labelled radiopharmaceuticals, skin exposure is mainly due to β-particles. This study aimed to investigate the equivalent dose saving of the staff when changing from an essentially manual radiolabelling procedure to an automatic dose dispenser (ADD). MATERIALS AND METHODS: The chemist and physician were asked to wear thermoluminescence dosimeters on their fingertips to evaluate the quantity of Hp(0.07) on the skin. Data collected were divided into two groups: before introducing ADD (no ADD) and after introducing ADD. RESULTS: For the chemist, the mean values (95th percentile) of Hp(0.07) for no ADD and ADD are 0.030 (0.099) and 0.019 (0.076) mSv/GBq, respectively, for Y, and 0.022 (0.037) and 0.007 (0.023) mSv/GBq, respectively, for Lu. The reduction for ADD was significant (t-test with P<0.05) for both isotopes. The relative differences before and after ADD collected for every finger were treated using the Wilcoxon test, proving a significantly higher reduction in extremity dose to each fingertip for Lu than for Y (P<0.05). For the medical staff, the mean values of Hp(0.07) (95th percentile) for no ADD and ADD are 0.021 (0.0762) and 0.0143 (0.0565) mSv/GBq, respectively, for Y, and 0.0011 (0.00196) and 0.0009 (0.00263) mSv/GBq, respectively, for Lu. The t-test provided a P-value less than 0.05 for both isotopes, making the difference between ADD and no ADD significant. CONCLUSION: ADD positively affects the dose saving of the chemist in handling both isotopes. For the medical staff not directly involved with the introduction of the ADD system, the analysis shows a learning curve of the workers over a 5-year period. Specific devices and procedures allow staff skin dose to be limited.

Fioroni, F., Grassi, E., Giorgia, C., Sara, R., Piccagli, V., Filice, A., et al. (2016). Skin dose saving of the staff in 90Y/177Lu peptide receptor radionuclide therapy with the automatic dose dispenser. NUCLEAR MEDICINE COMMUNICATIONS, 37(10), 1046-1052 [10.1097/MNM.0000000000000548].

Skin dose saving of the staff in 90Y/177Lu peptide receptor radionuclide therapy with the automatic dose dispenser

MOSTACCI, DOMIZIANO;
2016

Abstract

OBJECTIVE: When handling Y-labelled and Lu-labelled radiopharmaceuticals, skin exposure is mainly due to β-particles. This study aimed to investigate the equivalent dose saving of the staff when changing from an essentially manual radiolabelling procedure to an automatic dose dispenser (ADD). MATERIALS AND METHODS: The chemist and physician were asked to wear thermoluminescence dosimeters on their fingertips to evaluate the quantity of Hp(0.07) on the skin. Data collected were divided into two groups: before introducing ADD (no ADD) and after introducing ADD. RESULTS: For the chemist, the mean values (95th percentile) of Hp(0.07) for no ADD and ADD are 0.030 (0.099) and 0.019 (0.076) mSv/GBq, respectively, for Y, and 0.022 (0.037) and 0.007 (0.023) mSv/GBq, respectively, for Lu. The reduction for ADD was significant (t-test with P<0.05) for both isotopes. The relative differences before and after ADD collected for every finger were treated using the Wilcoxon test, proving a significantly higher reduction in extremity dose to each fingertip for Lu than for Y (P<0.05). For the medical staff, the mean values of Hp(0.07) (95th percentile) for no ADD and ADD are 0.021 (0.0762) and 0.0143 (0.0565) mSv/GBq, respectively, for Y, and 0.0011 (0.00196) and 0.0009 (0.00263) mSv/GBq, respectively, for Lu. The t-test provided a P-value less than 0.05 for both isotopes, making the difference between ADD and no ADD significant. CONCLUSION: ADD positively affects the dose saving of the chemist in handling both isotopes. For the medical staff not directly involved with the introduction of the ADD system, the analysis shows a learning curve of the workers over a 5-year period. Specific devices and procedures allow staff skin dose to be limited.
2016
Fioroni, F., Grassi, E., Giorgia, C., Sara, R., Piccagli, V., Filice, A., et al. (2016). Skin dose saving of the staff in 90Y/177Lu peptide receptor radionuclide therapy with the automatic dose dispenser. NUCLEAR MEDICINE COMMUNICATIONS, 37(10), 1046-1052 [10.1097/MNM.0000000000000548].
Fioroni, Federica; Grassi, Elisa; Giorgia, Cavatorta; Sara, Rubagotti; Piccagli, Vando; Filice, Angelina; Mostacci, Domiziano; Versari, Annibale; Iori...espandi
File in questo prodotto:
Eventuali allegati, non sono esposti

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11585/560954
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? 1
  • Scopus 5
  • ???jsp.display-item.citation.isi??? 5
social impact