We present a system for the acquisition of EEG signals based on active electrodes and implementing a Driving Right Leg circuit (DgRL). DgRL allows for single-ended amplification and analog-to-digital conversion, still guaranteeing a common mode rejection in excess of 110 dB. This allows the system to acquire high-quality EEG signals essentially removing network interference for both wet and dry-contact electrodes. The front-end amplification stage is integrated on the electrode, minimizing the system's sensitivity to electrode contact quality, cable movement and common mode interference. The A/D conversion stage can be either integrated in the remote back-end or placed on the head as well, allowing for an all-digital communication to the back-end. Noise integrated in the band from 0.5 to 100 Hz is comprised between 0.62 and 1.3 μV, depending on the configuration. Current consumption for the amplification and A/D conversion of one channel is 390 μA. Thanks to its low noise, the high level of interference suppression and its quick setup capabilities, the system is particularly suitable for use outside clinical environments, such as in home care, brain-computer interfaces or consumer-oriented applications.

EEG acquisition system based on active electrodes with common-mode interference suppression by Driving Right Leg circuit / Guermandi, Marco; Bigucci, Alessandro; Eleonora Franchi Scarselli; Guerrieri, Roberto. - ELETTRONICO. - 2015-:(2015), pp. 7319065.3169-7319065.3172. (Intervento presentato al convegno 37th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, EMBC 2015 tenutosi a MiCo Center, Milano Congressi Center, ita nel 2015) [10.1109/EMBC.2015.7319065].

EEG acquisition system based on active electrodes with common-mode interference suppression by Driving Right Leg circuit

GUERMANDI, MARCO;BIGUCCI, ALESSANDRO;FRANCHI SCARSELLI, ELEONORA;GUERRIERI, ROBERTO
2015

Abstract

We present a system for the acquisition of EEG signals based on active electrodes and implementing a Driving Right Leg circuit (DgRL). DgRL allows for single-ended amplification and analog-to-digital conversion, still guaranteeing a common mode rejection in excess of 110 dB. This allows the system to acquire high-quality EEG signals essentially removing network interference for both wet and dry-contact electrodes. The front-end amplification stage is integrated on the electrode, minimizing the system's sensitivity to electrode contact quality, cable movement and common mode interference. The A/D conversion stage can be either integrated in the remote back-end or placed on the head as well, allowing for an all-digital communication to the back-end. Noise integrated in the band from 0.5 to 100 Hz is comprised between 0.62 and 1.3 μV, depending on the configuration. Current consumption for the amplification and A/D conversion of one channel is 390 μA. Thanks to its low noise, the high level of interference suppression and its quick setup capabilities, the system is particularly suitable for use outside clinical environments, such as in home care, brain-computer interfaces or consumer-oriented applications.
2015
Proceedings of the Annual International Conference of the IEEE Engineering in Medicine and Biology Society, EMBS
3169
3172
EEG acquisition system based on active electrodes with common-mode interference suppression by Driving Right Leg circuit / Guermandi, Marco; Bigucci, Alessandro; Eleonora Franchi Scarselli; Guerrieri, Roberto. - ELETTRONICO. - 2015-:(2015), pp. 7319065.3169-7319065.3172. (Intervento presentato al convegno 37th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, EMBC 2015 tenutosi a MiCo Center, Milano Congressi Center, ita nel 2015) [10.1109/EMBC.2015.7319065].
Guermandi, Marco; Bigucci, Alessandro; Eleonora Franchi Scarselli; Guerrieri, Roberto
File in questo prodotto:
Eventuali allegati, non sono esposti

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11585/551065
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? 1
  • Scopus 6
  • ???jsp.display-item.citation.isi??? 5
social impact