We evaluated an EPID-based in-vivo dosimetry algorithm (IVD) for complex VMAT treatments in clinical routine. 19 consecutive patients with head-and-neck tumors and treated with Elekta VMAT technique using Simultaneous Integrated Boost strategy were enrolled. In-vivo tests were evaluated by means of (i) ratio R between daily in-vivo isocenter dose and planned dose and (ii) γ-analysis between EPID integral portal images in terms of percentage of points with γ-value smaller than one (γ%) and mean γ-values (γmean), using a global 3%-3 mm criteria. Alert criteria of ±5% for R ratio, γ% < 90% and γmean > 0.67 were chosen. A total of 350 transit EPID images were acquired during the treatment fractions. The overall mean R ratio was equal to 1.002 ± 0.019 (1 SD), with 95.9% of tests within ±5%. The 2D portal images of γ-analysis showed an overall γmean of 0.42 ± 0.16 with 93.3% of tests within alert criteria, and a mean γ% equal to 92.9 ± 5.1% with 85.9% of tests within alert criteria. Relevant discrepancies were observed in three patients: a set-up error was detected for one patient and two patients showed major anatomical variations (weight loss/tumor shrinkage) in the second half of treatment. The results are supplied in quasi real-time, with IVD tests displayed after only 1 minute from the end of arc delivery. This procedure was able to detect when delivery was inconsistent with the original plans, allowing physics and medical staff to promptly act in case of major deviations between measured and planned dose.

Initial clinical experience with Epid-based in-vivo dosimetry for VMAT treatments of head-and-neck tumors / Cilla, Savino; Meluccio, Daniela; Fidanzio, Andrea; Azario, Luigi; Ianiro, Anna; Macchia, Gabriella; Digesù, Cinzia; Deodato, Francesco; Valentini, Vincenzo; Morganti, ALESSIO GIUSEPPE; Piermattei, Angelo. - In: PHYSICA MEDICA. - ISSN 1120-1797. - ELETTRONICO. - 32:1(2016), pp. 52-58. [10.1016/j.ejmp.2015.09.007]

Initial clinical experience with Epid-based in-vivo dosimetry for VMAT treatments of head-and-neck tumors

MORGANTI, ALESSIO GIUSEPPE;
2016

Abstract

We evaluated an EPID-based in-vivo dosimetry algorithm (IVD) for complex VMAT treatments in clinical routine. 19 consecutive patients with head-and-neck tumors and treated with Elekta VMAT technique using Simultaneous Integrated Boost strategy were enrolled. In-vivo tests were evaluated by means of (i) ratio R between daily in-vivo isocenter dose and planned dose and (ii) γ-analysis between EPID integral portal images in terms of percentage of points with γ-value smaller than one (γ%) and mean γ-values (γmean), using a global 3%-3 mm criteria. Alert criteria of ±5% for R ratio, γ% < 90% and γmean > 0.67 were chosen. A total of 350 transit EPID images were acquired during the treatment fractions. The overall mean R ratio was equal to 1.002 ± 0.019 (1 SD), with 95.9% of tests within ±5%. The 2D portal images of γ-analysis showed an overall γmean of 0.42 ± 0.16 with 93.3% of tests within alert criteria, and a mean γ% equal to 92.9 ± 5.1% with 85.9% of tests within alert criteria. Relevant discrepancies were observed in three patients: a set-up error was detected for one patient and two patients showed major anatomical variations (weight loss/tumor shrinkage) in the second half of treatment. The results are supplied in quasi real-time, with IVD tests displayed after only 1 minute from the end of arc delivery. This procedure was able to detect when delivery was inconsistent with the original plans, allowing physics and medical staff to promptly act in case of major deviations between measured and planned dose.
2016
Initial clinical experience with Epid-based in-vivo dosimetry for VMAT treatments of head-and-neck tumors / Cilla, Savino; Meluccio, Daniela; Fidanzio, Andrea; Azario, Luigi; Ianiro, Anna; Macchia, Gabriella; Digesù, Cinzia; Deodato, Francesco; Valentini, Vincenzo; Morganti, ALESSIO GIUSEPPE; Piermattei, Angelo. - In: PHYSICA MEDICA. - ISSN 1120-1797. - ELETTRONICO. - 32:1(2016), pp. 52-58. [10.1016/j.ejmp.2015.09.007]
Cilla, Savino; Meluccio, Daniela; Fidanzio, Andrea; Azario, Luigi; Ianiro, Anna; Macchia, Gabriella; Digesù, Cinzia; Deodato, Francesco; Valentini, Vincenzo; Morganti, ALESSIO GIUSEPPE; Piermattei, Angelo
File in questo prodotto:
Eventuali allegati, non sono esposti

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11585/543523
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? 6
  • Scopus 35
  • ???jsp.display-item.citation.isi??? 33