PURPOSE: To evaluate clinical safety and effectiveness of a novel calcium-incorporated titanium implant surface (Xpeed, MegaGen Implant Co., Gyeongbuk, South Korea). MATERIALS AND METHODS: Sixty patients were randomised to receive one to six implants in the maxilla with either calcium-incorporated (Xpeed) or control resorbable blasted media (RBM) surfaces, according to a parallel group design at two centres. Implants were submerged and exposed at three different endpoints in equal groups of 20 patients, each at 12, 10 and 8 weeks, respectively. Within 2 weeks, implants were functionally loaded with provisional or definitive prostheses. Outcome measures were prosthesis failures, implant failures, any complications and peri-implant marginal bone level changes. RESULTS: Thirty patients received 45 calcium-incorporated implants and 30 patients received 42 control titanium implants. Three years after loading four patients dropped-out from the Xpeed group and one from the RBM group. No prosthesis or implant failures occurred. There were no statistically significant differences between the groups for complications (P = 0.91; difference in proportions = 0.79 %; 95% CI -0.71 to 2.29) and mean marginal bone level changes (P = 0.88; mean difference = -0.02 mm; 95% CI -0.26 to 0.22). CONCLUSIONS: Both implant surfaces provided good clinical results and no significant difference was found when comparing titanium implants with a nanostructured calcium-incorporated surface versus implants with RBM surfaces.

Safety and effectiveness of early loaded maxillary titanium implants with a novel nanostructured calcium-incorporated surface (Xpeed): 3-year results from a pilot multicenter randomised controlled trial / Felice, Pietro; Grusovin, Maria Gabriella; Barausse, Carlo; Grandi, Giovanni; Esposito, Marco. - In: EUROPEAN JOURNAL OF ORAL IMPLANTOLOGY. - ISSN 1756-2406. - STAMPA. - 8:(2015), pp. 245-254.

Safety and effectiveness of early loaded maxillary titanium implants with a novel nanostructured calcium-incorporated surface (Xpeed): 3-year results from a pilot multicenter randomised controlled trial.

FELICE, PIETRO;BARAUSSE, CARLO;
2015

Abstract

PURPOSE: To evaluate clinical safety and effectiveness of a novel calcium-incorporated titanium implant surface (Xpeed, MegaGen Implant Co., Gyeongbuk, South Korea). MATERIALS AND METHODS: Sixty patients were randomised to receive one to six implants in the maxilla with either calcium-incorporated (Xpeed) or control resorbable blasted media (RBM) surfaces, according to a parallel group design at two centres. Implants were submerged and exposed at three different endpoints in equal groups of 20 patients, each at 12, 10 and 8 weeks, respectively. Within 2 weeks, implants were functionally loaded with provisional or definitive prostheses. Outcome measures were prosthesis failures, implant failures, any complications and peri-implant marginal bone level changes. RESULTS: Thirty patients received 45 calcium-incorporated implants and 30 patients received 42 control titanium implants. Three years after loading four patients dropped-out from the Xpeed group and one from the RBM group. No prosthesis or implant failures occurred. There were no statistically significant differences between the groups for complications (P = 0.91; difference in proportions = 0.79 %; 95% CI -0.71 to 2.29) and mean marginal bone level changes (P = 0.88; mean difference = -0.02 mm; 95% CI -0.26 to 0.22). CONCLUSIONS: Both implant surfaces provided good clinical results and no significant difference was found when comparing titanium implants with a nanostructured calcium-incorporated surface versus implants with RBM surfaces.
2015
Safety and effectiveness of early loaded maxillary titanium implants with a novel nanostructured calcium-incorporated surface (Xpeed): 3-year results from a pilot multicenter randomised controlled trial / Felice, Pietro; Grusovin, Maria Gabriella; Barausse, Carlo; Grandi, Giovanni; Esposito, Marco. - In: EUROPEAN JOURNAL OF ORAL IMPLANTOLOGY. - ISSN 1756-2406. - STAMPA. - 8:(2015), pp. 245-254.
Felice, Pietro; Grusovin, Maria Gabriella; Barausse, Carlo; Grandi, Giovanni; Esposito, Marco
File in questo prodotto:
Eventuali allegati, non sono esposti

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11585/541728
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? 3
  • Scopus 10
  • ???jsp.display-item.citation.isi??? 9
social impact