We have developed a tomograph for single photon emission imaging (SPECT) of the breast for the detection of small size tumors. The SPECT is mounted on a ring that is rotating around the breast with the patient in prone position. The breast will be imaged by two opposing detector heads of approximately 5x15 cmxcm each, with a field of view about 13 cm wide. Each head is made up of one pixilated NaI crystal matrix coupled to three Hamamatsu H8500 PMTs. A “general purpose” lead collimator is positioned in front of the crystal. Detailed simulations have been made for the optimization and the evaluation of the detector performances. Monte Carlo results indicate that tumors of 8 mm diameter are detectable with a tumor/background uptake ratio of 5:1. The experimental characterization of the detector head is presented. The rotating ring is now being assembled.

S. Vecchio, N. Belcari, P. Bennati, M. Camarda, R. Campanini, M.N. Cinti, et al. (2007). A single photon emission computer tomograph for breast cancer imaging. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH. SECTION A, ACCELERATORS, SPECTROMETERS, DETECTORS AND ASSOCIATED EQUIPMENT, 581, 84-87 [10.1016/j.nima.2007.07.034].

A single photon emission computer tomograph for breast cancer imaging

CAMPANINI, RENATO;LANCONELLI, NICO;
2007

Abstract

We have developed a tomograph for single photon emission imaging (SPECT) of the breast for the detection of small size tumors. The SPECT is mounted on a ring that is rotating around the breast with the patient in prone position. The breast will be imaged by two opposing detector heads of approximately 5x15 cmxcm each, with a field of view about 13 cm wide. Each head is made up of one pixilated NaI crystal matrix coupled to three Hamamatsu H8500 PMTs. A “general purpose” lead collimator is positioned in front of the crystal. Detailed simulations have been made for the optimization and the evaluation of the detector performances. Monte Carlo results indicate that tumors of 8 mm diameter are detectable with a tumor/background uptake ratio of 5:1. The experimental characterization of the detector head is presented. The rotating ring is now being assembled.
2007
S. Vecchio, N. Belcari, P. Bennati, M. Camarda, R. Campanini, M.N. Cinti, et al. (2007). A single photon emission computer tomograph for breast cancer imaging. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH. SECTION A, ACCELERATORS, SPECTROMETERS, DETECTORS AND ASSOCIATED EQUIPMENT, 581, 84-87 [10.1016/j.nima.2007.07.034].
S. Vecchio; N. Belcari; P. Bennati; M. Camarda; R. Campanini; M.N. Cinti; A. Del Guerra; E. Iampieri; N. Lanconelli; R. Pani; L. Spontoni
File in questo prodotto:
Eventuali allegati, non sono esposti

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11585/49338
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 8
  • ???jsp.display-item.citation.isi??? 8
social impact