The design and realization of an optical sensor for measuring haematic pH during extracorporeal circulation is presented. It consists of a chemical sensing element in contact with the blood, an interrogation optical head to externally probe the sensing element and the front-end electronics to acquire and process the information of interest. The fluorescein O-methacrylate 97% is used as the indicator. The developed system has been tested in-vitro and on an in-vivo animal model. It showed a linear behavior in the haematic range of interest with a mean error lower than 0.01 units of pH.
Ferrari, L., Fabbri, P., Luigi, R., Francesco, P. (2012). Haematic pH sensor for extracorporeal circulation. NY : SPIE - the international society for optics and photonics [10.1117/12.906562].
Haematic pH sensor for extracorporeal circulation
FABBRI, PAOLA;
2012
Abstract
The design and realization of an optical sensor for measuring haematic pH during extracorporeal circulation is presented. It consists of a chemical sensing element in contact with the blood, an interrogation optical head to externally probe the sensing element and the front-end electronics to acquire and process the information of interest. The fluorescein O-methacrylate 97% is used as the indicator. The developed system has been tested in-vitro and on an in-vivo animal model. It showed a linear behavior in the haematic range of interest with a mean error lower than 0.01 units of pH.I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.