Carbon dioxide angiography is based on the visualization (i.e., the radiographic contrast) of gas bubbles injected in blood vessels. By using an experimental X-ray bench, the energy response of a flat panel detector has been measured (Varian CB4030) and, with a dedicated phantom and a software simulation, the image contrast of vessels is injected with Iodine and CO2. Moreover, the dynamical behavior of a moving gas bubble has been studied with the software simulator. The results show that the contrast generated by carbon dioxide is about one fourth of that obtained with iodine, demonstrating that CO2 angiography should use different radiological settings with respect to iodine angiography. In particular, a kVp increase has a lower reduction of contrast-to-noise-ratio (CNR) with carbon dioxide than with iodinated contrast medium (CM), suggesting possible technological improvements both on radiological emission and image enhancement methods.

DAVID BIANCHINI, PIER LUCA ROSSI, GIACOMO FELICIANI, ALESSANDRO LOMBI, IVAN CORAZZA, ROMANO ZANNOLI (2015). CARBON DIOXIDE ANGIOGRAPHY: SIMULATION OF OPERATIVE CONDITIONS FOR DIAGNOSTIC IMAGE OPTIMIZATION. JOURNAL OF MECHANICS IN MEDICINE AND BIOLOGY, 15, 1540023-1-1540023-8 [10.1142/S0219519415400230].

CARBON DIOXIDE ANGIOGRAPHY: SIMULATION OF OPERATIVE CONDITIONS FOR DIAGNOSTIC IMAGE OPTIMIZATION

BIANCHINI, DAVID;ROSSI, PIER LUCA;FELICIANI, GIACOMO;LOMBI, ALESSANDRO;CORAZZA, IVAN;ZANNOLI, ROMANO
2015

Abstract

Carbon dioxide angiography is based on the visualization (i.e., the radiographic contrast) of gas bubbles injected in blood vessels. By using an experimental X-ray bench, the energy response of a flat panel detector has been measured (Varian CB4030) and, with a dedicated phantom and a software simulation, the image contrast of vessels is injected with Iodine and CO2. Moreover, the dynamical behavior of a moving gas bubble has been studied with the software simulator. The results show that the contrast generated by carbon dioxide is about one fourth of that obtained with iodine, demonstrating that CO2 angiography should use different radiological settings with respect to iodine angiography. In particular, a kVp increase has a lower reduction of contrast-to-noise-ratio (CNR) with carbon dioxide than with iodinated contrast medium (CM), suggesting possible technological improvements both on radiological emission and image enhancement methods.
2015
DAVID BIANCHINI, PIER LUCA ROSSI, GIACOMO FELICIANI, ALESSANDRO LOMBI, IVAN CORAZZA, ROMANO ZANNOLI (2015). CARBON DIOXIDE ANGIOGRAPHY: SIMULATION OF OPERATIVE CONDITIONS FOR DIAGNOSTIC IMAGE OPTIMIZATION. JOURNAL OF MECHANICS IN MEDICINE AND BIOLOGY, 15, 1540023-1-1540023-8 [10.1142/S0219519415400230].
DAVID BIANCHINI;PIER LUCA ROSSI;GIACOMO FELICIANI;ALESSANDRO LOMBI;IVAN CORAZZA;ROMANO ZANNOLI
File in questo prodotto:
Eventuali allegati, non sono esposti

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11585/463770
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 4
  • ???jsp.display-item.citation.isi??? 4
social impact