AutoCapture based on the evoked response can be confounded by electrode polarization. In this study, polarization was measured in human subjects who had chronic atrial leads. The aim of the study was to determine whether electrode polarization can be measured using a time integral atrial evoked-response integral (AERI) of the negative portion of the atrial paced ER evoked-response signal and to determine whether high-polarization atrial leads unsuitable for AutoCapture can be identified a priori. Atrial intracardiac-electrogram (IEGM) signals from 39 patients with implanted pacemakers were recorded and analyzed. The signals were recorded during conventional atrial-threshold searches. A total of 221 atrial-capture thresholds were recorded, ranging from 0.25 to 2.75 V with a mean of 0.79 V. Each evoked response was evaluated using the AERI in a 36 ms window following the 0.4 ms atrial stimulus. The polarization was estimated as a linear function of stimulus voltage using the evoked-response signal integral of captured beats identified on the IEGM. The 221 threshold-search datasets were obtained using leads with eight different electrode materials. Polarization could be measured using AERI as a function of stimulus voltage. Furthermore, this polarization measure can be used to identify high-polarization leads, which are ill suited for the atrial AutoCapture algorithm.

Evaluation of the atrial evoked response for capture detection with high-polarization leads / Sperzel J; Binner L; Boriani G; Biffi M; Snell J; Scheiner J; Park E; Bornzin G. - In: PACING AND CLINICAL ELECTROPHYSIOLOGY. - ISSN 0147-8389. - STAMPA. - 28:(2005), pp. s57-s62. [10.1111/j.1540-8159.2005.00059.x]

Evaluation of the atrial evoked response for capture detection with high-polarization leads.

BORIANI, GIUSEPPE;BIFFI, MAURO;
2005

Abstract

AutoCapture based on the evoked response can be confounded by electrode polarization. In this study, polarization was measured in human subjects who had chronic atrial leads. The aim of the study was to determine whether electrode polarization can be measured using a time integral atrial evoked-response integral (AERI) of the negative portion of the atrial paced ER evoked-response signal and to determine whether high-polarization atrial leads unsuitable for AutoCapture can be identified a priori. Atrial intracardiac-electrogram (IEGM) signals from 39 patients with implanted pacemakers were recorded and analyzed. The signals were recorded during conventional atrial-threshold searches. A total of 221 atrial-capture thresholds were recorded, ranging from 0.25 to 2.75 V with a mean of 0.79 V. Each evoked response was evaluated using the AERI in a 36 ms window following the 0.4 ms atrial stimulus. The polarization was estimated as a linear function of stimulus voltage using the evoked-response signal integral of captured beats identified on the IEGM. The 221 threshold-search datasets were obtained using leads with eight different electrode materials. Polarization could be measured using AERI as a function of stimulus voltage. Furthermore, this polarization measure can be used to identify high-polarization leads, which are ill suited for the atrial AutoCapture algorithm.
2005
Evaluation of the atrial evoked response for capture detection with high-polarization leads / Sperzel J; Binner L; Boriani G; Biffi M; Snell J; Scheiner J; Park E; Bornzin G. - In: PACING AND CLINICAL ELECTROPHYSIOLOGY. - ISSN 0147-8389. - STAMPA. - 28:(2005), pp. s57-s62. [10.1111/j.1540-8159.2005.00059.x]
Sperzel J; Binner L; Boriani G; Biffi M; Snell J; Scheiner J; Park E; Bornzin G
File in questo prodotto:
Eventuali allegati, non sono esposti

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11585/39510
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 8
  • ???jsp.display-item.citation.isi??? 2
social impact