A HPLC method with UV detection has been developed for the simultaneous determination of levomepromazine, clozapine and their main metabolites: N-desmethyl-levomepromazine, levomepromazine sulfoxide, O-desmethyl-levomepromazine, N-desmethylclozapine and clozapine N-oxide. The analytes were separated on a C8 reversed-phase column using a mobile phase composed of acetonitrile and a pH 2.0, 34 mM phosphate buffer containing 0.3% triethylamine (29:71, v/v). Loxapine was used as the internal standard. A reliable biological sample pre-treatment procedure by means of solid-phase extraction on C1 cartridges was implemented, which allows to obtain good extraction yields (>91%) for all analytes and appropriate sample purification from endogenous interference. The method was validated in terms of extraction yield, precision and accuracy. These assays gave R.S.D.% values for precision always lower than 4.9% and mean accuracy values higher than 93%. The method is suitable for the therapeutic drug monitoring (TDM) of patients undergoing polypharmacy with levomepromazine and clozapine.

Simultaneous determination of the antipsychotic drugs levomepromazine and clozapine and their main metabolites in human plasma by a HPLC-UV method

MERCOLINI, LAURA;RAGGI, MARIA AUGUSTA
2007

Abstract

A HPLC method with UV detection has been developed for the simultaneous determination of levomepromazine, clozapine and their main metabolites: N-desmethyl-levomepromazine, levomepromazine sulfoxide, O-desmethyl-levomepromazine, N-desmethylclozapine and clozapine N-oxide. The analytes were separated on a C8 reversed-phase column using a mobile phase composed of acetonitrile and a pH 2.0, 34 mM phosphate buffer containing 0.3% triethylamine (29:71, v/v). Loxapine was used as the internal standard. A reliable biological sample pre-treatment procedure by means of solid-phase extraction on C1 cartridges was implemented, which allows to obtain good extraction yields (>91%) for all analytes and appropriate sample purification from endogenous interference. The method was validated in terms of extraction yield, precision and accuracy. These assays gave R.S.D.% values for precision always lower than 4.9% and mean accuracy values higher than 93%. The method is suitable for the therapeutic drug monitoring (TDM) of patients undergoing polypharmacy with levomepromazine and clozapine.
2007
L. Mercolini; F. Bugamelli; E. Kenndler; G. Boncompagni; L. Franchini; M.A. Raggi
File in questo prodotto:
Eventuali allegati, non sono esposti

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11585/35110
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? 3
  • Scopus 48
  • ???jsp.display-item.citation.isi??? 44
social impact