An HPLC method has been developed for the simultaneous determination of aromatic and terpenic constituents of cloves on a C8 RP column, with the mobile phase consisting of a pH 3.5 phosphate buffer-triethylamine (30%) and acetonitrile (70%); a flow rate of 1.2 mL/min and a diode-array detector were used. Complete separation of all analytes (eugenol (EUG), eugenol acetate (AEUG), beta-caryophyllene, alpha-humulene and caryophyllene oxide) was achieved within 7 min. Good linearity was found in the range 0.125–40.0 µg/mL for EUG and AEUG and in the range 0.250–20.0 µg/mL for the terpenic compounds. After validation, the method was successfully applied to the analysis of clove oil and clove extract samples. The results obtained indicate good accuracy (recovery percentage mean value corresponding to 99.9%) and satisfactory precision.
A. Musenga, A. Ferranti, M.A. Saracino, S. Fanali, M.A. Raggi (2006). Simultaneous determination of aromatic and terpenic constituents of cloves by means of HPLC with diode array detection. JOURNAL OF SEPARATION SCIENCE, 29, 1251-1258 [10.1002/jssc.200600023].
Simultaneous determination of aromatic and terpenic constituents of cloves by means of HPLC with diode array detection
MUSENGA, ALESSANDRO;FERRANTI, ANNA;SARACINO, MARIA ADDOLORATA;RAGGI, MARIA AUGUSTA
2006
Abstract
An HPLC method has been developed for the simultaneous determination of aromatic and terpenic constituents of cloves on a C8 RP column, with the mobile phase consisting of a pH 3.5 phosphate buffer-triethylamine (30%) and acetonitrile (70%); a flow rate of 1.2 mL/min and a diode-array detector were used. Complete separation of all analytes (eugenol (EUG), eugenol acetate (AEUG), beta-caryophyllene, alpha-humulene and caryophyllene oxide) was achieved within 7 min. Good linearity was found in the range 0.125–40.0 µg/mL for EUG and AEUG and in the range 0.250–20.0 µg/mL for the terpenic compounds. After validation, the method was successfully applied to the analysis of clove oil and clove extract samples. The results obtained indicate good accuracy (recovery percentage mean value corresponding to 99.9%) and satisfactory precision.I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.