Purpose: The aim of this study is to assess and compare the accuracy of two different patient-specific instrumentation (PSI) systems for total knee replacement, both intra-operatively for bone preparation and post-operatively for final component alignment. Methods: Twenty-five patients were treated according to a computer tomography (CT)-based PSI system (group A) and 25 to a magnetic resonance imaging (MRI)/X-ray-based system (group B). Alignments on the three anatomical planes and resection thickness at the cutting blocks and at the resulting bone cuts were recorded intra-operatively by a standard surgical navigation system. Alignments of the prosthetic components and mechanical axis were also measured post-operatively on radiographs. These measurements at both the femur and tibia were compared with those of the corresponding pre-operative planning, considering discrepancies larger than 3° as outliers. Results: In both groups, the mean absolute differences between pre-operatively planned alignments and corresponding intra- and post-operative measurements ranged from a minimum of 1.2° to a maximum of 2.9° in all three anatomical planes. In both groups and in both femur and tibia, the plane with the smallest percentage of outliers was the coronal, maximum 17 %. The comparison between two groups was statistically significant (p = 0.02) in the femoral sagittal plane, where group B showed smaller alignment discrepancies at the cutting blocks. Conclusions: Both PSI systems showed good alignments in the coronal plane in all stages. For a few measurements, a better performance was observed in the MRI/X-ray-based system than in the CT-based system. Level of evidence: I.

Ensini A, Timoncini A, Cenni F, Belvedere C, Fusai F, Leardini A, et al. (2014). Intra and post-operative accuracy assessments of two different patient-specific instrumentatio systems for total knee replacement. KNEE SURGERY, SPORTS TRAUMATOLOGY, ARTHROSCOPY, 22, 621-629 [10.1007/s00167-013-2667-9].

Intra and post-operative accuracy assessments of two different patient-specific instrumentatio systems for total knee replacement.

ENSINI, ANDREA;TIMONCINI, ANTONIO;GIANNINI, SANDRO
2014

Abstract

Purpose: The aim of this study is to assess and compare the accuracy of two different patient-specific instrumentation (PSI) systems for total knee replacement, both intra-operatively for bone preparation and post-operatively for final component alignment. Methods: Twenty-five patients were treated according to a computer tomography (CT)-based PSI system (group A) and 25 to a magnetic resonance imaging (MRI)/X-ray-based system (group B). Alignments on the three anatomical planes and resection thickness at the cutting blocks and at the resulting bone cuts were recorded intra-operatively by a standard surgical navigation system. Alignments of the prosthetic components and mechanical axis were also measured post-operatively on radiographs. These measurements at both the femur and tibia were compared with those of the corresponding pre-operative planning, considering discrepancies larger than 3° as outliers. Results: In both groups, the mean absolute differences between pre-operatively planned alignments and corresponding intra- and post-operative measurements ranged from a minimum of 1.2° to a maximum of 2.9° in all three anatomical planes. In both groups and in both femur and tibia, the plane with the smallest percentage of outliers was the coronal, maximum 17 %. The comparison between two groups was statistically significant (p = 0.02) in the femoral sagittal plane, where group B showed smaller alignment discrepancies at the cutting blocks. Conclusions: Both PSI systems showed good alignments in the coronal plane in all stages. For a few measurements, a better performance was observed in the MRI/X-ray-based system than in the CT-based system. Level of evidence: I.
2014
Ensini A, Timoncini A, Cenni F, Belvedere C, Fusai F, Leardini A, et al. (2014). Intra and post-operative accuracy assessments of two different patient-specific instrumentatio systems for total knee replacement. KNEE SURGERY, SPORTS TRAUMATOLOGY, ARTHROSCOPY, 22, 621-629 [10.1007/s00167-013-2667-9].
Ensini A; Timoncini A; Cenni F; Belvedere C; Fusai F; Leardini A; Giannini S.
File in questo prodotto:
Eventuali allegati, non sono esposti

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11585/265930
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? 11
  • Scopus 44
  • ???jsp.display-item.citation.isi??? 41
social impact