BACKGROUND: Many ICD carriers experience inappropriate shocks, but the relative merits of dual- /single-chamber devices for arrhythmia discrimination still remain unclear. We explored possible advantages of the atrial data provided by dual-chamber implantable defibrillators (ICD) for discrimination of real-life supraventricular/ventricular tachyarrhythmias (SVT/VT). METHODS: 100 dual-chamber traces from 24 ICD were blindly reviewed in dual-chamber and simulated single-chamber (with/without discriminator data) reading modes by five electrophysiologists who determined chamber of origin and provided Likert-scale "confidence" ratings. We assessed 1) intra/interobserver concordance; 2) diagnostic accuracy, using expert diagnoses as a reference standard; 3) ROC curves of sensitivity/specificity of "likelihood perception" scores, generated by combining chamber-of-origin diagnostic judgments with Likert-scale "confidence" ratings. We also assessed diagnostic accuracy of automated discrimination by all possible dual-/single-chamber algorithm configurations. RESULTS: Interobserver concordance was "substantial" (modified Cohen kappa-test values for dual-/single-chamber, 0.79/0.68); intraobserver concordance "almost complete" (kappa ≥ 0.89). Dual-chamber mode provided best diagnostic sensitivity/specificity (99%/92%) and highest reader confidence (p<0.001). Area under ROC curves of sensitivity/specificity values for the "likelihood perception" score (representing electrophysiologists' perceptions of the likelihood that an episode was of ventricular origin) was highest in dual-chamber mode (0.98 vs. 0.93 for both single-chamber modes; p<0.001). Regarding automated discrimination, all four dual-chamber configurations conferred 100% sensitivity (specificity values ranged 39%-88%), whereas single-chamber configurations appeared inferior (best sensitivity/specificity combination, 89%/64%). CONCLUSIONS: Availability of the atrial channel helps in reducing inappropriate ICD therapies by providing relevant advantages in terms of both appropriate cardiologist's post-hoc discrimination of SVT/VT (improving program tailoring) and automated arrhythmia discrimination.
Titolo: | Arrhythmia discrimination by physician and defibrillator: importance of atrial channel. | |
Autore/i: | DIEMBERGER, IGOR; MARTIGNANI, CRISTIAN; BIFFI, MAURO; FRABETTI, LORENZO; VALZANIA, CINZIA; Cooke RM; RAPEZZI, CLAUDIO; BRANZI, ANGELO; BORIANI, GIUSEPPE | |
Autore/i Unibo: | ||
Anno: | 2012 | |
Rivista: | ||
Digital Object Identifier (DOI): | http://dx.doi.org/10.1016/j.ijcard.2010.09.005 | |
Abstract: | BACKGROUND: Many ICD carriers experience inappropriate shocks, but the relative merits of dual- /single-chamber devices for arrhythmia discrimination still remain unclear. We explored possible advantages of the atrial data provided by dual-chamber implantable defibrillators (ICD) for discrimination of real-life supraventricular/ventricular tachyarrhythmias (SVT/VT). METHODS: 100 dual-chamber traces from 24 ICD were blindly reviewed in dual-chamber and simulated single-chamber (with/without discriminator data) reading modes by five electrophysiologists who determined chamber of origin and provided Likert-scale "confidence" ratings. We assessed 1) intra/interobserver concordance; 2) diagnostic accuracy, using expert diagnoses as a reference standard; 3) ROC curves of sensitivity/specificity of "likelihood perception" scores, generated by combining chamber-of-origin diagnostic judgments with Likert-scale "confidence" ratings. We also assessed diagnostic accuracy of automated discrimination by all possible dual-/single-chamber algorithm configurations. RESULTS: Interobserver concordance was "substantial" (modified Cohen kappa-test values for dual-/single-chamber, 0.79/0.68); intraobserver concordance "almost complete" (kappa ≥ 0.89). Dual-chamber mode provided best diagnostic sensitivity/specificity (99%/92%) and highest reader confidence (p<0.001). Area under ROC curves of sensitivity/specificity values for the "likelihood perception" score (representing electrophysiologists' perceptions of the likelihood that an episode was of ventricular origin) was highest in dual-chamber mode (0.98 vs. 0.93 for both single-chamber modes; p<0.001). Regarding automated discrimination, all four dual-chamber configurations conferred 100% sensitivity (specificity values ranged 39%-88%), whereas single-chamber configurations appeared inferior (best sensitivity/specificity combination, 89%/64%). CONCLUSIONS: Availability of the atrial channel helps in reducing inappropriate ICD therapies by providing relevant advantages in terms of both appropriate cardiologist's post-hoc discrimination of SVT/VT (improving program tailoring) and automated arrhythmia discrimination. | |
Data prodotto definitivo in UGOV: | 2013-06-19 10:53:53 | |
Appare nelle tipologie: | 1.01 Articolo in rivista |