The comparison of two HPLC methods, one with electrochemical detection and the other with coulometric detection, for the simultaneous analysis of catecholamines and 4-hydroxy-3-methoxyphenylethylene glycol (MHPG) in human plasma is presented. The careful pretreatment of plasma samples is based on an innovative two-step procedure by means of solid phase extraction (SPE) which uses one single hydrophilic-lipophilic balance cartridge. The extraction yield values found were higher than 85% for epinephrine, norepinephrine and MHPG, and higher than 70% for dopamine. The assays carried out on real plasma samples with the coulometric system gave good results in terms of sensitivity (limits of quantitation: 0.10-0.15 ng/ml for catecholamines, 0.6 ng/ml for MHPG) and selectivity, while interference was sometimes found when using the amperometric system. Precision was also satisfactory, with relative standard deviation values for intermediate precision always lower than 6%. The HPLC method with coulometric detection coupled to a novel SPE procedure is thus suitable for the simultaneous determination of catecholamines and MHPG in plasma of volunteers subjected to experimental stress.
C. Sabbioni, M.A. Saracino, R. Mandrioli, S. Pinzauti, S. Furlanetto, G. Gerra, et al. (2004). Simultaneous liquid chromatographic analysis of catecholamines and 4-hydroxy-3-methoxyphenylethylene glycol (MHPG) in human plasma: comparison of amperometric and coulometric detection. JOURNAL OF CHROMATOGRAPHY A, 1032, 65-71 [10.1016/j.chroma.2004.01.008].
Simultaneous liquid chromatographic analysis of catecholamines and 4-hydroxy-3-methoxyphenylethylene glycol (MHPG) in human plasma: comparison of amperometric and coulometric detection
SABBIONI, CESARE;SARACINO, MARIA ADDOLORATA;MANDRIOLI, ROBERTO;RAGGI, MARIA AUGUSTA
2004
Abstract
The comparison of two HPLC methods, one with electrochemical detection and the other with coulometric detection, for the simultaneous analysis of catecholamines and 4-hydroxy-3-methoxyphenylethylene glycol (MHPG) in human plasma is presented. The careful pretreatment of plasma samples is based on an innovative two-step procedure by means of solid phase extraction (SPE) which uses one single hydrophilic-lipophilic balance cartridge. The extraction yield values found were higher than 85% for epinephrine, norepinephrine and MHPG, and higher than 70% for dopamine. The assays carried out on real plasma samples with the coulometric system gave good results in terms of sensitivity (limits of quantitation: 0.10-0.15 ng/ml for catecholamines, 0.6 ng/ml for MHPG) and selectivity, while interference was sometimes found when using the amperometric system. Precision was also satisfactory, with relative standard deviation values for intermediate precision always lower than 6%. The HPLC method with coulometric detection coupled to a novel SPE procedure is thus suitable for the simultaneous determination of catecholamines and MHPG in plasma of volunteers subjected to experimental stress.I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.