Although spectrofluorimetry is a very sensitive analytical technique, its use is limited by the non linear relationship between the concentration of the analyte of interest and the generated electric signal. Here we discuss a very easy correction method which takes into account all the instrumental factors and transforms the observed intensity value in a quantity, the corrected luminescence intensity, that is directly proportional to the concentration of the observed luminophore. At the end of the discussion, a general formula is presented together with two examples illustrating the method and its application. © 1998 Elsevier Science B.V.
Credi, A., Prodi, L. (1998). From observed to corrected luminescence intensity of solution systems: An easy-to-apply correction method for standard spectrofluorimeters. SPECTROCHIMICA ACTA. PART A, MOLECULAR AND BIOMOLECULAR SPECTROSCOPY, 54(1), 159-170 [10.1016/S1386-1425(97)00224-2].
From observed to corrected luminescence intensity of solution systems: An easy-to-apply correction method for standard spectrofluorimeters
Credi A.;Prodi L.
1998
Abstract
Although spectrofluorimetry is a very sensitive analytical technique, its use is limited by the non linear relationship between the concentration of the analyte of interest and the generated electric signal. Here we discuss a very easy correction method which takes into account all the instrumental factors and transforms the observed intensity value in a quantity, the corrected luminescence intensity, that is directly proportional to the concentration of the observed luminophore. At the end of the discussion, a general formula is presented together with two examples illustrating the method and its application. © 1998 Elsevier Science B.V.I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.