Abstract: A signal processing-based procedure is proposed for calibrating an experimental sensor-based test system used to identify material model parameters. A standard dynamic mechanical analyzer (DMA) sensorized test apparatus is considered, enabling the measurement of dynamic excitation and displacement response in a specimen under flexural conditions. To account for the dynamic contributions of the system frame and fixtures to the measured response, a novel calibration procedure is introduced, mainly differing from the techniques used in standard test applications. A multi-degree-of-freedom dynamic model of the instrument frame, coupled with the beam specimen under test, is considered, and a frame identification procedure is described. The proposed procedure requires measurements from at least three beam specimens made of a known material but with different geometries. It is shown that an accurate frame model can be identified using an algebraic numerical technique. It is shown that the accuracy of the material model identification can be improved by applying the proposed calibration technique. Some experimental application examples are presented and discussed.

Catania, G., Amadori, S. (2025). Dynamical Calibration of a Standard Sensor-Based Experimental Test Apparatus for Accurate Material Model Parametric Identification. SENSORS, 25(7), 1-27 [10.3390/s25071982].

Dynamical Calibration of a Standard Sensor-Based Experimental Test Apparatus for Accurate Material Model Parametric Identification

Catania, Giuseppe
Primo
;
Amadori, Stefano
Secondo
2025

Abstract

Abstract: A signal processing-based procedure is proposed for calibrating an experimental sensor-based test system used to identify material model parameters. A standard dynamic mechanical analyzer (DMA) sensorized test apparatus is considered, enabling the measurement of dynamic excitation and displacement response in a specimen under flexural conditions. To account for the dynamic contributions of the system frame and fixtures to the measured response, a novel calibration procedure is introduced, mainly differing from the techniques used in standard test applications. A multi-degree-of-freedom dynamic model of the instrument frame, coupled with the beam specimen under test, is considered, and a frame identification procedure is described. The proposed procedure requires measurements from at least three beam specimens made of a known material but with different geometries. It is shown that an accurate frame model can be identified using an algebraic numerical technique. It is shown that the accuracy of the material model identification can be improved by applying the proposed calibration technique. Some experimental application examples are presented and discussed.
2025
Catania, G., Amadori, S. (2025). Dynamical Calibration of a Standard Sensor-Based Experimental Test Apparatus for Accurate Material Model Parametric Identification. SENSORS, 25(7), 1-27 [10.3390/s25071982].
Catania, Giuseppe; Amadori, Stefano
File in questo prodotto:
File Dimensione Formato  
sensors-25-01982.pdf

accesso aperto

Tipo: Versione (PDF) editoriale / Version Of Record
Licenza: Licenza per Accesso Aperto. Creative Commons Attribuzione (CCBY)
Dimensione 8.63 MB
Formato Adobe PDF
8.63 MB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11585/1009715
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 0
  • ???jsp.display-item.citation.isi??? 0
social impact