A volumetric microsampling (VAMS) device (20 μl) was evaluated and validated for the analysis of γ-hydroxybutyric acid (GHB) in venous blood using a simple ultra-high-pressure liquid chromatography-tandem mass spectrometry (UHPLC-MS/MS) method. GHB was extracted from VAMS device by acetonitrile, after a re-hydration step in a temperature-controlled ultrasonic bath at 60°C for 10 min. Chromatographic analysis was carried out on a Kinetex C18 column using 0.1% formic acid in water and acetonitrile as binary gradient mobile phase (from 5 to 95% of acetonitrile from 1 to 2.5 min) at a flow rate of 0.3 ml/min. The VAMS method was fully validated according to current guidelines with satisfactory results in terms of linearity, selectivity, precision, absolute recovery, matrix effect and stability. The linearity was determined from 0.5 to 200 μg/ml and the lower limit of quantitation was 0.5 μg/ml. The novel VAMS-UHPLC-MS/MS method was successfully compared with plasma-based method in a GHB-treated patient as a proof of concept.

Development and validation of volumetric absorptive microsampling coupled with UHPLC-MS/MS for the analysis of gamma-hydroxybutyric acid in human blood

Riva, Roberto;Plazzi, Giuseppe;Contin, Manuela
2020

Abstract

A volumetric microsampling (VAMS) device (20 μl) was evaluated and validated for the analysis of γ-hydroxybutyric acid (GHB) in venous blood using a simple ultra-high-pressure liquid chromatography-tandem mass spectrometry (UHPLC-MS/MS) method. GHB was extracted from VAMS device by acetonitrile, after a re-hydration step in a temperature-controlled ultrasonic bath at 60°C for 10 min. Chromatographic analysis was carried out on a Kinetex C18 column using 0.1% formic acid in water and acetonitrile as binary gradient mobile phase (from 5 to 95% of acetonitrile from 1 to 2.5 min) at a flow rate of 0.3 ml/min. The VAMS method was fully validated according to current guidelines with satisfactory results in terms of linearity, selectivity, precision, absolute recovery, matrix effect and stability. The linearity was determined from 0.5 to 200 μg/ml and the lower limit of quantitation was 0.5 μg/ml. The novel VAMS-UHPLC-MS/MS method was successfully compared with plasma-based method in a GHB-treated patient as a proof of concept.
2020
Mohamed, Susan; Riva, Roberto; Moresco, Monica; Plazzi, Giuseppe; Contin, Manuela
File in questo prodotto:
Eventuali allegati, non sono esposti

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11585/752901
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? 0
  • Scopus 5
  • ???jsp.display-item.citation.isi??? 4
social impact