The aim of this study was to carry out a bench evaluation of the biomechanical feasibility of carbon dioxide (CO2) coronary arteriography. Many patients among the aging population of individuals requiring cardiac intervention have underlying renal insufficiency making them susceptible to contrast-induced nephropathy. To include those patients, it is imperative to find an alternative and safe technique to perform coronary imaging on cardiac ischemic patients. As CO2angiography has no renal toxicity, it may be a possible solution offering good imaging with negligible collateral effects. Theoretically, by carefully controlling the gas injection process, new automatic injectors may avoid gas reflux into the aorta and possible cerebral damage. A feasibility study is mandatory. A mechanical mock of the coronary circulation was developed and employed. CO2was injected into the coronary ostium with 2 catheters (2F and 6F) and optical images of bubbles flowing inside the vessels at different injection pressures were recorded. The gas behavior was then carefully studied for quantitative and qualitative analysis. Video recordings showed that CO2injection at a precise pressure in the interval between the arterial dicrotic notch and the minimum diastolic value does not result in gas reflow into the aorta. Gas reflow was easier to control with the smaller catheter, but the gas bubbles were smaller with different vascular filling. Our simulation demonstrates that carefully selected injection parameters allow CO2coronary imaging without any risk of gas reflux into the aorta.

Carbon dioxide coronary angiography: A mechanical feasibility study with a cardiovascular simulator / Corazza, Ivan*; Taglieri, Nevio; Pirazzini, Edoardo; Rossi, Pier Luca; Lombi, Alessandro; Scalise, Filippo; Caridi, James G.; Zannoli, Romano. - In: AIP ADVANCES. - ISSN 2158-3226. - ELETTRONICO. - 8:1(2018), pp. 015225.1-015225.11. [10.1063/1.5016601]

Carbon dioxide coronary angiography: A mechanical feasibility study with a cardiovascular simulator

Corazza, Ivan
;
Taglieri, Nevio;Rossi, Pier Luca;Lombi, Alessandro;Zannoli, Romano
2018

Abstract

The aim of this study was to carry out a bench evaluation of the biomechanical feasibility of carbon dioxide (CO2) coronary arteriography. Many patients among the aging population of individuals requiring cardiac intervention have underlying renal insufficiency making them susceptible to contrast-induced nephropathy. To include those patients, it is imperative to find an alternative and safe technique to perform coronary imaging on cardiac ischemic patients. As CO2angiography has no renal toxicity, it may be a possible solution offering good imaging with negligible collateral effects. Theoretically, by carefully controlling the gas injection process, new automatic injectors may avoid gas reflux into the aorta and possible cerebral damage. A feasibility study is mandatory. A mechanical mock of the coronary circulation was developed and employed. CO2was injected into the coronary ostium with 2 catheters (2F and 6F) and optical images of bubbles flowing inside the vessels at different injection pressures were recorded. The gas behavior was then carefully studied for quantitative and qualitative analysis. Video recordings showed that CO2injection at a precise pressure in the interval between the arterial dicrotic notch and the minimum diastolic value does not result in gas reflow into the aorta. Gas reflow was easier to control with the smaller catheter, but the gas bubbles were smaller with different vascular filling. Our simulation demonstrates that carefully selected injection parameters allow CO2coronary imaging without any risk of gas reflux into the aorta.
2018
Carbon dioxide coronary angiography: A mechanical feasibility study with a cardiovascular simulator / Corazza, Ivan*; Taglieri, Nevio; Pirazzini, Edoardo; Rossi, Pier Luca; Lombi, Alessandro; Scalise, Filippo; Caridi, James G.; Zannoli, Romano. - In: AIP ADVANCES. - ISSN 2158-3226. - ELETTRONICO. - 8:1(2018), pp. 015225.1-015225.11. [10.1063/1.5016601]
Corazza, Ivan*; Taglieri, Nevio; Pirazzini, Edoardo; Rossi, Pier Luca; Lombi, Alessandro; Scalise, Filippo; Caridi, James G.; Zannoli, Romano
File in questo prodotto:
File Dimensione Formato  
1.5016601.pdf

accesso aperto

Tipo: Versione (PDF) editoriale
Licenza: Licenza per Accesso Aperto. Creative Commons Attribuzione (CCBY)
Dimensione 3.87 MB
Formato Adobe PDF
3.87 MB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11585/621198
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 3
  • ???jsp.display-item.citation.isi??? 4
social impact